RIGOL
Chapter 5 Troubleshooting & Maintenance
Service Guide for DM3068
5-6
5.2.3 Components Inspection
In this part, you will get more information concerning the fault location for removing
advanced malfunction.
1.
Fuse
The DM3068 has three types of fuse, which are: power fuse, current input fuse
on the rear panel and internal current input fuse on the mainboard, the
parameters are as follows:
115V Power fuse: AC, 250V, T250mA, slow- melt, 5x20mm
230V Power fuse: AC, 250V, T125mA, slow- melt, 5x20mm
Replaceable 200mA current input fuse on the rear panel: AC, 250V, F500mA,
quick-melt, 5x20mm
Internal 10A current input fuse on the mainboard: AC, 250V, F10A, quick-melt,
6x32mm
Inspection method
:
Take out the fuse and execute continuity test using the
meter. For the internal 10A current input fuse on the mainboard: separately
insert the test lead ends into socket HI and 10A and then switch the meter to
continuity test; short-circuit the test lead ends, if the testing result is connecting,
the fuse is normal, otherwise it may have been blown out.
2.
Power Frequency Transformer
The DM3068 is powered by a power frequency transformer. The power socket is
equipped with a fuse socket and switch. You can select an input winding for the
transformer through the voltage selector in order to accommodate different
mains voltage. For countries using 220V voltage as the mains voltage, 230 V is
necessary and a T125mA fuse should be installed to protect the meter; while
115V is recommended and a T250mA fuse should be installed if 110 V mains
voltage is supplying, otherwise the meter could not be started.
Inspection procedures:
(1)
Cut off the power supply.
(2)
Setup the voltage selector.
(3)
Remove the case.
(4)
Re-power on and keep the power switch in “1”.
(5)
Check the voltage of every winding of transformer.
(6)
Cur off the power supply again.
(7)
Remove the fault and fit up the case after re-inspection.
WARNING
Keep all body parts away from the conductors inside the instrument when
the meter is powered without case, otherwise you may get electrocuted!
!