Service Guide for DM3068
2-5
2. 2
Testing Equipments
The following table lists the equipments recommended for the test. If you do not
have such equipments, use others that have the same Precisions instead.
Table 2-2 Recommended equipments
Test and
Calibrate Items
Recommended
Equipments
[1]
Precision
Requirements
Zero Test
None
Pure copper 4-terminal
short-circuiter
DC Voltage
Fluke 5520A and Agilent
3458A (contain accessory 002) <1/5 dmm 24 hour spec.
DC Current
Fluke 5520A and Agilent
3458A (contain accessory 002) <1/5 dmm 24 hour spec.
Resistance
Fluke 5520A and Agilent
3458A (contain accessory 002) <1/5 dmm 24 hour spec.
AC Voltage
Fluke 5520A
<1/5 dmm 24 hour spec.
AC Current
Fluke 5520A
<1/5 dmm 24 hour spec.
Frequency
Fluke 5520A
<1/5 dmm 24 hour spec.
Capacitance
Fluke 5520A
<1/5 dmm 24 hour spec.
Remark
[1]
: Agilent 3458A (contain accessory 002) characterizes Fluke 5520A.