User’s Guide for DG2000 Series
2-50
To Save the Instrument State
Users are allowed to store the instrument state in any one of the 4 Non-Volatile
Memories. When the power is on again, the instrument will return to the state before it
was powered off. The state storage will “memorize” the selected function (including
the Arbitrary Waveform), Frequency, Amplitude, DC Offset, Duty Cycle, Symmetry,
and other modulation parameters used.
To Save the Instrument State, following the steps:
1.
Choose the file Type to store
Press Store/Recall
Æ
State , and set the storage type as “State”.
2.
Choose the location of the file.
There are four positions in the local STATE1, STATE2, STATE3 and STATE4, and
you can choose one of them by rotating the knob.
3.
Name the file and Save it
Press Store button, enter the self-defined name. Press Store to finish.