RIGOL
© 2006 RIGOL Technologies, Inc
User’s Guide for DG2000 Series
4-11
Square Wave period upper limit= 1Ms
The Square Wave Period upper limit is 1Ms. If the specified period is greater than 1Ms,
the system will adjust the period to 1Ms automatically.
Square Wave period lower limit= 25ns
The Square Wave Period lower limit is 25ns. If the specified period is less than 25ns,
the system will adjust the period to 25ns automatically.
Ramp Wave period upper limit= 1Ms
。
The Ramp Wave Period upper limit is 1Ms. If the specified period is greater than 1Ms,
the system will adjust the period to 1Ms automatically.
Ramp Wave period lower limit= 2.5
μ
s
。
The Ramp Wave Period lower limit is 2.5ns. If the specified period is less than 2.5ns,
the system will adjust the period to 2.5ns automatically.
Pulse Wave period upper limit= 2Ks
。
The Pulse Wave Period upper limit is 2Ks. If the specified period is greater than 2Ks,
the system will adjust the period to 2Ks automatically.
Pulse Wave period lower limit= 62.5ns
。
The Pulse Wave Period lower limit is 62.5ns. If the specified period is less than 62.5ns,
the system will adjust the period to 62.5ns automatically.
Arb Wave period upper limit= 1Ms
。
The Arb Wave Period upper limit is 1Ms. If the specified period is greater than 1Ms,
the system will adjust the period to 1Ms automatically.
Arb Wave period lower limit= 83.3ns
。
The Arb Wave Period lower limit is 83.3ns. If the specified period is less than 83.3ns,
the system will adjust the period to 83.3ns automatically.
Amplitude upper limit= **
Different mode has different amplitude upper limit; amplitude upper limit may vary in
the same mode if the load is different.
Amplitude lower limit= **
Different mode has different amplitude lower limit; amplitude lower limit may vary in
the same mode if the load is different.
High level upper limit = **
Different mode has different High Level upper limit; High Level upper limit may vary in
the same mode if the load is different.
Low level lower negative limit=**
Different mode has different High Level lower limit; High Level lower limit may vary in
the same mode if the load is different.
High level must be greater than Low level.
High level must always be greater than Low level. If the high level is set less than the