Rigel SafeTest 60 Instruction Manual
–
Revision 1.2
30
Appendix A Pass / Fail Limits of IEC 60601-1
Earthbond test limit at 25A, 50Hz
Excluding power cord
< 0.1 Ω
Including power cord
< 0.2 Ω
Type B
Applied Parts
Type BF
Applied Parts
Type CF
Applied Parts
Leakage Current Type
NC
SFC
NC
SFC
NC
SFC
Earth Leakage (3rd edition)*
500
0μA 10000μA 5000μA 10000μA 5000μA 10000μA
Earth Leakage (General)
500μA
1000μA
500μA
1000μA
500μA
1000μA
Enclosure Leakage
100μA
500μA
100μA
500μA
100μA
500μA
Patient Leakage (dc)
10μA
50μA
10μA
50μA
10μA
50μA
Patient Leakage (ac)
100μA
500μA
100μA
500μA
10μA
50μA
Patient Leakage (F-Type)
NA
NA
NA
5000μA
NA
50μA
Patient
Leakage
(Mains
on
SIP/SOP)
NA
5mA
NA
NA
NA
NA
Patient Auxiliary Current (dc)
10μA
50μA
10μA
50μA
10μA
50μA
Patient Auxiliary Current (ac)
100μA
500μA
100μA
500μA
10μA
50μA
* The pass fail limit for Earth Leakage in the 3rd edition of IEC 60601 has been increased from
500
μA under normal condition to 5000μA for class I equipment with NO exposed metal parts that
may become live when a fault appears.