Rigel SafeTest 60 Instruction Manual
–
Revision 1.2
22
9 Insulation testing
To perform an insulation test, press
from HOME screen
The SafeTest 60 will automatically start the test and will automatically stop by
pressing the EARTHBOND, LEAKAGE, INSULATION, POINT TO POINT or
HOME icon.
During this test, 500V D.C. is applied between the black socket on the back
panel as well as earth pin in the EUT socket and both the live and neutral
pins of the EUT.
The INSULATION VOLTAGE will be applied between L-N to EARTH for class 1
equipment or L-N to the black socket for class 2 equipment.
Connection between EUT and SafeTest 60;
INSULATION TESTING IEC 62353
Do not exceed the maximum permitted voltage of 30 V AC/DC with respect
to earth potential! Electric Shock danger!
Do not connect any probe combination to voltages in excess of 30 V
AC/DC with respect to earth potential when performing non-power
tests. This can cause damage to the equipment.
Insulation resistan
ce limit MΩ
– Class I Equipment
– Class II Equipment
>2MΩ
>7MΩ