17
6 NFPA-99 LEAKAGE testing
To perform an NFPA-99 leakage test, press
from HOME screen to enter
LEAKAGE test. If the required test standard needs to change, please press
HOME and see 2.1.
To apply mains voltage in NORMAL POLARITY and START the test, press the
button on the front panel. The test will run until the
button is pressed.
To apply mains in REVERSED POLARITY and START the test, press the
button on the front panel. The test will run until the
button is pressed.
To minimise the power breaks during your tests, please see 8.
will appear on the screen when the DUT socket is activated.
Connection between EUT and SafeTest 60;
GROUND LEAKAGE NFPA-99
M
ai
ns
in
M
ai
ns