Rigel SafeTest 60 Instruction Manual
–
Revision 1.2
16
will appear on the screen when the DUT socket is activated.
When the SafeTest 60 is set to test to IEC 62353, the SINGLE FAULT buttons
and
are deactivated in order to perform the test as per IEC 62353
requirements.
Connection between EUT and SafeTest 60;
EQUIPMENT LEAKAGE IEC 62353
Current in µA (RMS)
ALL
APPLIED
PART
Equipment leakage
– direct method.
– Class I Equipment
– Class II Equipment (touch current)
0.5mA
0.1mA
M
ai
ns
in
M
ai
ns