(10) Adjust the position of cryogenic gas suction tube to direct the flow toward the sample center, 

as illustrated in Fig.11. From the view point at the front of the equipment, adjust the center of 
the cryogenic gas suction pipe to be aligned with the sample.   

 

2.3    Evaluate operation and performance 

(1)  Click 

New Sample

 under 

File

 in the 

CrystalClear

 menu bar. 

(2)  Initialize Instrument and then, set up a measurement according to the following conditions for 

one image within flow-bar step 

Initial Images

 

Start Angle (˚)  End Angle (˚) 

Width (˚) 

Exposure time (sec) 

30 

31 

1.0 

16 

 

(3)  Inspect the image and make sure there is not a shadow of the suction pipe in it. If it has a 

shadow of the suction pipe, you will see something similar to the image in Fig.12. In such a 
case, adjust the height of the suction pipe. 

 

(4)  Close 

CrystalClea

r. Turn off the power for all components. Shut down the system.

 

Shadow of the 

suction pipe 

Fig.12 Checking image 

 

ON
 

OF

Cooling gas 

alignment guide 

Sample pin 

Sample center 

Fig.11 Adjust the position of cryogenic gas suction 

Содержание XtaLAB mini

Страница 1: ...Cat No 2366A201 XtaLAB mini Cryogenic gas suction attachment Instruction manual Manual No ME12973A02...

Страница 2: ...ual has to be purchased for each product If there are any missing or incorrectly collated pages in the delivered instruction manual please contact the sales representative who you purchased the instru...

Страница 3: ...view 1 2 Cryogenic gas suction attachment installation and adjustment 2 2 1 Wiring the electric power supply cable 2 2 2 Fine alignment of the cryogenic gas suction pipe 5 2 3 Evaluate operation and p...

Страница 4: ......

Страница 5: ...cold gas flow and reduce the amount of ice accumulation Please refer to or read other manuals including XtaLAB mini system Maintenance Manual 11568C and CrystalClear Ver 2 0 0 for CCD system for Smal...

Страница 6: ...alog window to get synchronization between goniometer and software 5 Move the Omega axis to 90 degrees 6 Close the CrystalClear software and turn off power to the XtaLAB mini and the water chiller sys...

Страница 7: ...be 12 Confirm the cable at the rear side of goniometer goes through the hole along the Omega axis Pass the cable through the wiring duct Fig 5 near the USB ports at the rear of the XtaLAB mini 13 Remo...

Страница 8: ...the suction pipe when switching on the cryogenic gas suction at the left Mount the pipe such that the height from the top to the main body is around 56 mm Fig 8 Fix the suction pipe into the main bod...

Страница 9: ...and beam stop 2 2 Fine alignment of the cryogenic gas suction pipe 1 Turn on the switch on the main body of cryogenic gas suction 2 Turn on the power of XtaLAB mini and the water chiller system 3 Lau...

Страница 10: ...according to the following conditions for one image within flow bar step Initial Images Start Angle End Angle Width Exposure time sec 30 31 1 0 16 3 Inspect the image and make sure there is not a sha...

Страница 11: ...tion Rigaku Americas Corporation Rigaku Europe SE Head Office 3 9 12 Matsubara cho Akishima shi Tokyo 196 8666 Japan Phone 81 42 545 8189 Fax 81 42 544 9223 e mail rinttyo rigaku co jp 9009 New Trails...

Отзывы: