Tapes geometry measurement system
[Version 1.0] 08
th
December 2014
11
10.3.2. Thermometers calibration
stop the calibration process by pressing
Stop
button
open the calibration menu
Settings > Calibration
(Fig. 12)
choose the number of temperature sensor (
Sensor index
field
)
direct the sensor to the surface with stated temperature, having entered the
temperature value in
Temp. 1
or
Temp. 2
field and press
Set
button. When the
operation is finished successfully the message
Calibration parameter has
been successfully set
appears. Detailed information on temperature sensors
can be found in OMEGA OS136-1-MA documentation
perform the operation described in the previous paragraph for two different
temperatures. By pressing the Set button the previous calibration value is over-
written
Figure 12
10.4. Settings
All sensors in the system are numbered (Index) from 1 to 18, where the 1-6
match the sensors 1-6 from width measurement group of sensors, numbers 7-15 in the
Settings match the sensors 1-9 from width measurement group of sensors and numbers
16-18 match the three temperature sensors. Indexes T1-T6 indicates the results of indi-
rect thickness measurement.
To perform the system settings enter the General menu.
in the
Timeout
field the delay in milliseconds between successive cycles of
the sensors polling is set
to set the tolerances choose the number of the sensor in the
Index
field and
set the upper and lower limits in the fields
Upper limit
and
Lower limit
corre-
spondingly
to set the production task number use the field
Task Number