Tapes geometry measurement system
[Version 1.0] 08
th
December 2014
10
Figure 10
10.3. Calibration
10.3.1. Calibration of the thickness measurement micrometers
Calibration procedure:
remove tapes from the shaft
start measurement process by pressing
«
Start
»
open the calibration menu:
Settings > Set etalon/Sw. State
(Fig. 11)
select the numbers of the micrometers to be calibrated (numbers 7-15 corre-
spond to 9 micrometers of the tape thickness control subsystem)
make sure that
Etalons
mode is selected and
Set
option in
Mode
field is on
than press
Apply
button
Figure 11