A.3. Adjustment
A.3.1. Preparation
1. Remove the ADF exposure glass and the main exposure glass.
2. Remove the top shield plate.
3. Connect the optical adjustment harness [A] to CN11 on the MBU board.
4. Set the ADF exposure glass [B] and the optical adjustment glass [C] as
shown above.
5. Set the resolution chart [D] on the optical adjustment glass so that the
leading edge of the chart is against the ADF exposure glass.
NOTE: Make sure that the end of the test chart with the thin white line
pattern is on the edge [E] of the glass that is nearest the ADF
exposure glass.
6. Loosen the four screws on the SBU.
7. Set dip switches 1, 7, and 8 to the ON position.
8. Set the SCSI rotary switch to 1.
9. Turn on the power.
I
CAUTION:
Cover the transformer with some paper, to avoid electric shocks.
10. Press the reset switch to put the lump on.
[A]
G404X516.wmf
[B]
[C]
[D]
[E]
G404X514.wmf
APPENDIX A. OPTICAL ADJUSTMENT (MECHANICAL)
October 9th, 1995
Adjustment
A-2