Guidance and manufacturer’s declaration - electromagnetic immunity (cont’d)
The HIVADOT® is intended for use in the electromagnetic environment specified
below. The customer or the user of HIVADOT® should assure that it is used in such
an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic
environment guidance
Power
frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency mag-
netic fields should be at
levels characteristic of a
typical location in a typical
commercial or hospital
environment.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.