![Richmar HIVADOT Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/richmar/hivadot/hivadot_instruction-manual_1454543021.webp)
21
OPERATING INSTRUCTIONS
1. PREPARING BEFORE THERAPY
Charging the Device
The HIVADOT® device will come with a partial charge. In order to
complete a treatment, it is recommended to charge the device once
it decreases to lower than a 30% charge.
Place the Type C end of the charging cable into the top of the device
port by sliding the cover over the two output ports to cover them and
expose the charging port.
Place the other end of the charging cable into the ac adapter box and
then plug the ac adapter box into the electrical outlet on the wall. The
device will show a charging icon on the screen, in the upper right hand
corner.
NOTE:
The device cannot be used while charging.
NOTE:
1. If the rechargeable battery of the devices is not fully charged
when you start a treatment, the batteries may run out during
the treatment. We advise you to
ALWAYS
fully charge the
device before you start a treatment. A completely depleted
battery should take approximately 2 hours to fully charge.
Medical Device
2.
ALWAYS
disconnect the power adapter from the wall
and the device after charging is completed.
3. While the battery is charging, the device does not work.
Treatment Preparation
The HIVADOT® device requires an absolutely dry surface for
treatment. Before treatment, prepare the patients skin by making
sure there is no moisture in the treatment area. If needed, baby
powder can be used in the area to help soak up any moisture.