Richmar HIVADOT Скачать руководство пользователя страница 26

26

Each Favorite has a 3 Stage default. To delete a phase: 

1.  Press [

] to highlight the last stage. 

2.  Press [

] and it will highlight the HZ parameter only,  

 in  blue. 

3.  Press the [

] to highlight the trashcan in blue.  

4.  Press the [

] to confirm deletion of that stage. 

Custom Protocol Setup

STEP 1: Select a Favorite

1.  From the main menu, Custom Protocols should be highlighted 

blue. Press the [

] to advance to the next screen. 

2.  Select one from Favorites 1-5 by pressing the up [

] or 

down [

] arrow and then press [

] to choose the desired 

Favorite.

*To re-add a stage, that only has 2 stages, simply highlight the  
parameters and press the [

] to highlight the + sign in the upper 

right side of the screen and if you want to change the parameters of 

the added stage, advance to the Parameter Changes Section of the 

manual.  

NOTE:

 The + sign will only be present on the device if there 

are less than 3 stages.

Содержание HIVADOT

Страница 1: ...DEEP OSCILLATION THERAPY DEVICE INSTRUCTION MANUAL A...

Страница 2: ...h Brands does not guarantee its contents and reserves the right to improve and amend it at any time without prior notice Amendments may however be published in new editions of this manual All Right Re...

Страница 3: ...Device and Accessories__________________________ 17 Operating Instructions_____________________________________ 21 1 Preparing Before Therapy______________________________ 21 2 Start Treatment________...

Страница 4: ...Health Brands Richmar a division of Compass Health Brands reserves the right to change the software and associated data without notice All other rights reserved For questions about your HIVADOT Deep...

Страница 5: ...an authorized repair facility Any modifications or repairs performed by unauthorized centers or groups will void this warranty RICHMAR SHALL RESERVE THE RIGHT TO REQUEST PROOF OF PURCHASE FROM THE END...

Страница 6: ...ation Any representative or agreement not contained in the warranty shall be void Information in this document is subject to change without prior notice The manufacturer of the equipment may have pate...

Страница 7: ...please read this manual thoroughly and become familiar with the controls as well as the accessories before operating the unit OPERATOR QUALIFICATIONS The HIVADOT is intended for use by medical special...

Страница 8: ...Device This side up The transportation package must be vertical and straight up during transportation Medical Device Fragile handle with care The product inside the packaging could be easily damaged i...

Страница 9: ...have the potential to cause minor to moderate injury or damage to equipment Medical Device WARNING Text with a WARNING indicator will explain possible safety infractions that will potentially cause se...

Страница 10: ...DO NOT use the device in the presence of a flammable anesthetic mixture with air or oxygen or nitrous oxide In case of device failure or other obvious defects switch the unit off immediately and noti...

Страница 11: ...taneously with other therapeutic device such as microwave shortwave to avoid mis operation Operation in close proximity e g 3 ft to a shortwave or microwave therapy device may produce instability in t...

Страница 12: ...iagnosed heart disease should follow precautions recommended by their physicians before using this device Keep this device out of the reach of children DO NOT use this device while sleeping The main d...

Страница 13: ...astic case Ensure that the connectors are fully dry before connecting to the device Use this device only with the accessories recommended by the manufacturer Handle the device with care DO NOT drop kn...

Страница 14: ...ins 5 Edema reduction 6 Improvement of wound healing CONTRAINDICATIONS Patients with the following disease diseases or conditions MUST NOT use the HIVADOT 1 Acute infections 2 Acute inflammations with...

Страница 15: ...mpedance 6 M Timer Max 90 mins divided into up to three stages each stage is 1 30 mins default 20 mins stepping 1 min ELECTRICAL Power Supply of Adapter Input AC100 240V 50 60Hz output DC 5V 2A Batter...

Страница 16: ...llowing No Description Item QTY UOM 1 The HIVADOT device DO1009 1 each 2 HIVADOT stand DO1009STND 1 each 3 Wall Adapter USB Power Cord DO1009X 1 each 4 Pin Lead Wires for Electrodes White DO1009PINLW...

Страница 17: ...2 3 Type C power adapter port 4 Display screen 5 Power Button and Select Button 6 UP button 7 DOWN button 8 START PAUSE SKIP Phase button 9 Back STOP button NOTE Both output ports on the top of the d...

Страница 18: ...18 LEAD WIRES Pin Connection Lead wires for Electrodes White 2 0mm plug 2 pk Item DO1009PINLW Applicator Titanium Element Lead Wire Grey 4 0mm plug 2 pk Item DO1009APPLW...

Страница 19: ...ATING HEADS 1 Detachable oscillating heads 2 Applicator Handle 3 Applicator Connection 4 Applicator Titanium Element Lead Wire Connector ADHESIVE ELECTRODE 10cm Oscillating Head 5 5cm Oscillating Head...

Страница 20: ...20 TITANIUM NEUTRAL ELEMENT VINYL GLOVES 100 bx HIVADOT STAND NOTE Use replacement parts only supplied by the manufacturer Contact your local dealer to place an order...

Страница 21: ...per right hand corner NOTE The device cannot be used while charging NOTE 1 If the rechargeable battery of the devices is not fully charged when you start a treatment the batteries may run out during t...

Страница 22: ...and It is recommended for the patient to either place the titanium element between the web of two fingers or hold it with a light grip Medical Device WARNING Use this device only with the accessories...

Страница 23: ...including lifting edges The adhesive electrodes and membranes are for single patient use only DO NOT use on another patient to prevent infection If the adhesive electrodes loses viscosity please repl...

Страница 24: ...sen requires connection to an electrode it must connect with a WHITE lead wire How to Apply Therapy The electrostatic pulse occurs between opposing polarities by way of the titanium rod or electrode f...

Страница 25: ...t Button 1 Press and hold button to turn on or turn off the device 2 Press to select setting Up Button 1 Press to go up to desired option 2 Increase adjustable parameters including intensity Down Butt...

Страница 26: ...cols should be highlighted blue Press the to advance to the next screen 2 Select one from Favorites 1 5 by pressing the up or down arrow and then press to choose the desired Favorite To re add a stage...

Страница 27: ...ge the parameter 3 Press and to adjust to the desired Output Frequency Hz 4 To accept press again to highlight solid blue again Adjust Modulation and Timer 1 Press to highlight the parameter blue you...

Страница 28: ...lue There are 4 options to choose from Athletic Rehab Chiropractic Continuing Care 2 Press the button to select The only paramater that is adjustable in Preset Protocols is the timer 3 To adjust the t...

Страница 29: ...patient know it is advancing to the next stage The device will do this for each stage change NOTE To skip a phase press and hold the to skip the current phase and move to the next before the timer is...

Страница 30: ...art treatment when ready Stopping Treatment 1 The device will either emit a beeping sound when all three phases are complete and treatment is done 2 OR you may press and hold the in the left corner an...

Страница 31: ...isinfectant wipe with 70 or less medical isopropanol alcohol Manufacturer suggests using Dismozon plus or similar product Oscillating heads will likely need replacement after approximately 25 30 uses...

Страница 32: ...ed at least monthly to ensure proper operation of this device 1 Power adapter Check to make sure the cord is not frayed kinked and does not have torn or cut insulation 2 Connection cables Check to mak...

Страница 33: ...the connector 2 Connect new battery and wire to the device wire making sure it is inserted like a puzzle piece in the right direction Place the battery back in the compartment so it is flush against...

Страница 34: ...egulations If you have any questions please contact the local authorities responsible for waste disposal This symbol indicates that batteries contain substances that may be harmful to human health and...

Страница 35: ...on Preparing Before Therapy The battery is worn out Replace with a new battery see section Battery Replacement The device cannot be charged The adapter is not connected properly Check to ensure the ad...

Страница 36: ...s is high Use of the HIVADOT adjacent to or stacked with other equipment should be avoided because it could result in improper operation If such use is necessary this equipment and the other equipment...

Страница 37: ...1 The HIVADOT use RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Cl...

Страница 38: ...for power supply lines Main power quality should be that of a commercial or hospital environment 1 kV for input output lines Surge IEC 61000 4 5 0 5kV 1 kV line s to lines 0 5kV 1 kV line s to lines...

Страница 39: ...e that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Power frequency 50 60Hz magnetic field IEC 61000 4 8 30 A m 30 A m Pow...

Страница 40: ...lectromagnetic environment guidance Conducted RF 3 Vrms 150 kHz 80 MHz 6 V rms in ISM bands between 150 kHz and 80 MHz 80 AM at 1 kHz 3 Vrms 150 kHz 80 MHz 6 V rms in ISM bands between 150 kHz and 80...

Страница 41: ...to 470 GMRS 460 FRS 460 FM c 5 Hz deviation 1 kHz sine 2 0 3 28 710 745 780 704 to 787 LTE Band 13 17 Pulse Modulation b 217 Hz 0 2 0 3 9 810 870 930 800 to 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 85...

Страница 42: ...ST LEVEL V m 5240 5500 5785 5100 to 5800 WLAN 802 11 a n Pulse Modulation b 217 Hz 0 2 0 3 9 NOTE a For some services only the uplink frequencies are included b The carrier shall be modulated using a...

Страница 43: ...y an authori zed repair technician approved by the manufacturer For assistance on ordering replacement parts or technical support please contact Richmar a division of Compass Health Brands Ph 888 549...

Страница 44: ...44...

Страница 45: ...45...

Страница 46: ...46...

Страница 47: ...47...

Страница 48: ...Manufactured for 3541 22 04 A 2022 Compass Health Brands Corp Made in China...

Отзывы: