16
Calibrating the Novo-Gloss Flex 60
The Rhopoint Novo-Gloss Flex 60 is supplied with 2 calibration standards for the GU &
GUh ranges. The standards are magnetically enclosed which protects the glass
reference tiles from contamination and damage. The holder can be opened for
inspection and cleaning.
The calibration tiles are enclosed in a magnetically sealed holder.
To calibrate the instrument, remove the measurement adapter and attach the
calibration holder. The instrument serial number on the calibration holder and
instrument must be matched.
Place the measurement head and calibration holder on a flat surface and initiate calibration.
Select the correct calibration range in the calibration menu - the range is indicated on
labels on the top and bottom of each holder. If the incorrect range is selected for the
attached tile Calibration Error will be shown on the screen. Cancel the calibration and
select the correct range then restart the procedure.