RHE26/27 Installation & Startup Guide – Doc. No. 8.2.1.06 – Ver. 1.40
23
Examples:
a)
Configuration of pulse output [R, S] for mass:
o
Select value 1 for parameter ID [R/S01]
o
Acc. to Table 10, the parameter [R/S02] has to be selected. This parameter specifies
the mass equivalent to one pulse. E.g. when set to 0.001kg, a mass transfer of 1kg
will cause 1000 pulses
b)
Configuration of a frequency output [R, S] for mass flow rate:
o
Select value 13 for parameter [R/S01]
o
Acc. to Table 10 the parameters [R/S04 & 07] have to be selected. The parameter
[R/S04] specifies the mass flow rate equivalent to the frequency set under
parameter [R/S07]. E.g. when [R/S04] is set to 1kg/min and [R/S07] is set to 1000Hz
a mass flow rate of 5kg/min will cause a frequency of 5000Hz
c)
Configuration of a digital output [I-L] for mass flow alarm:
o
Select value 1 for parameter [I-L01]
o
Acc. to Table 10 the parameters [I-L02, 03 & 04] apply. Select the needed alarm type
[I-L02] and enter the low [I-L03] and high [I-L04] alarm values. The unit of the alarm
is determined by the digital output configuration [I-L01]. For mass flow alarm the
unit is fixed to kg/min
Table 4: Important Pulse and Digital Output Parameters
ID
Abbreviation
Full Name / Description
[R/S01]
VersDigOutCh0/1Config
Versatile Digital Output Configuration:
Used to assign different functions to the pulse output.
In simple pulse mode, Pulse1 and Pulse2 operate independently. In phase shift
mode, Pulse1 and Pulse2 work together to produce the desired pulse output
relationship. In phase shift mode, both outputs (Pulse1 & Pulse2) should be set to
the same mode.
0: Pulse output is off.
1: Mass Flow – Simple Pulse, Forward Flow. → set [R/S02]
2: Volume Flow – Simple Pulse, Forward Flow. → set [R/S03]
3: Mass Flow – Pulse1 & Pulse2 90 Degrees phase shift. → set [R/S02]
4: Volume Flow – Pulse1 & Pulse2 90 Degrees phase shift. → set [R/S03]
5: Mass Flow – Simple Pulse, Reverse Flow. → set [R/S02]
6: Volume Flow – Simple Pulse, Reverse Flow. → set [R/S03]
7: Mass Flow Rate Fast Frequency Output. → set [R/S04 & 07] fast
8: Volumetric Flow Rate Fast Frequency Output. → set [R/S05 & 07] response
9: Density Fast Frequency Output. → set [R/S06 & 07] time
10: Mass Flow Rate Slow Frequency Output. → set [R/S04 & 07] slow
11: Volumetric Flow Rate Slow Frequency Output. → set [R/S05 & 07] response
12: Density Slow Frequency Output. → set [R/S06 & 07] time (1s)
13: Mass Flow Rate RHE Compatibility Mode Frequency Output.
→ set [R/S04 & 07] Freq. mode of
14: Volume Flow Rate RHE Compatibility Mode Frequency Output. previous RHE
→ set [R/S05 & 07]
Содержание RHE26
Страница 28: ...RHE26 27 Installation Startup Guide Doc No 8 2 1 06 Ver 1 40 28 7 Ordering Code 7 1 RHE26 7 2 RHE27 ...
Страница 40: ...Appendix B Technical Information Ver 1 00 40 B 2 2 RHE27 RHE27 E1 Transmitter Dimensions all dimensions in mm ...
Страница 41: ...Appendix B Technical Information Ver 1 00 41 RHE27 E2 Transmitter Dimensions all dimensions in mm ...
Страница 42: ...Appendix C CE Certificates Ver 1 00 42 Appendix C CE Certificates ...
Страница 43: ...Appendix C CE Certificates Ver 1 00 43 ...
Страница 44: ...Document No 8 2 1 06 Version 1 40 APR 2019 ...