Page 8 of 22
REACTORS
This manual refers to the following UV disinfections systems.
Table 1a
Model No
Series
Length
No. Reactors in unit
Diameter mm
940001
4000-D-28 (LCE 4)
895mm
2
62
940002
15000-S-170 (LCS 17)
1210mm
1
76
Table 1b
Model No
Series
Inlet/outlet Ø
Min. space to service
reactor (mm)
Max. flow rate
L/min
940001
LCE 4 - 4000-D-28
2” BSP Nipples
1000
83
*
940002
LCS 17 - 15000-S-170
2” BSP Nipples
1300
250
*
*Maximum flow rate is at UV dose
25mJ/cm
2
. If in doubt a water sample must be taken and tested. Refer to
Applications on page 3.
INSTALLATION STANDARDS
The UV system must be installed:
by a qualified person, and
in accordance the installation instructions, and
in compliance with Standards AS/NZS 3500.4, AS/NZS 3000 and all local codes and regulatory
authority requirements.
In New Zealand the installation must also conform with the New Zealand Building Code.
All packaging materials must be removed from the warm water and UV unit prior to their installation.
WARM WATER UNIT APPLICATION
This UV system is designed for the purpose of disinfecting potable water. Its use in an application
other than this may shorten its life or reduce its effectiveness.
If this UV system is to be used where an uninterrupted warm water supply is necessary for your
application or business you should ensure that you have back up redundancy within the warm water
system design. This should ensure the continuity of the warm water system in the event that this UV
system were to become inoperable for any reason. We recommend you seek advice from your
plumber or specifier about your needs and building back up redundancy into your warm water system.