29 - ElS
( For technicians only ): Draining solenoid valve shut. (15)
30 - PoF
Machine status message. Pressure less than 0.8 bar. (1)
31 - Pon
Machine status message. Pressure higher than 0.8 bar. (1)
32 - S.L
Machine status message. Sufficient water in tank.
33 - APE
Machine status message. Door open.
34 - noL
Machine status message. Water below minimum level.
35 - CH.
Door shut.
36 - SCr
For technicians only.
37 - LEt
For technicians only.
38 - SI
Makes sure that the microprocessor is not faulty.
39 - Err
Microprocessor error
40 - - - -
On ( electronic board powered ).
41 -ACU
Maximum vacuum test.
42 - drY
Supplementary drying program to select after the sterilizing cycle has terminated.
43 - oCH
Vacuum OK message. There are no leaks in the hydraulic circuit of the machine.
44 - LLn
Vacuum faulty message. There are leaks in the hydraulic circuit of the machine.
45 - TEST
VUOTO OK
Vacuum test message
.
46 - USd
Message indicating dirty water in machine.
47 - FrA
Fractionated vacuum with 3 peaks before sterilizing
48 - ElP
( For technicians only ): Vacuum pump N. C. solenoid
There is also an intermittent acoustic signal along with the fault messages. The machine will
block.
RESET
: The machine can
be reset and made operative by pressing one of the 4 buttons or by powering the board again. The
board will, however, abort the cycle and repeat the operations from the beginning (the door must be
opened).
The
dor
message inhibits the board functions until the boiler door is opened.
Warning: if the machine is powered by the electricity main,
electricians who check or test the electrical system must NEVER
disconnect the power actuator (triac) for any reason whatsoever.