Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast
Rev. 09 of 18-04-02 - E:\Documenti1\Manuali superati non in uso\Vacuum Star GB-R09.doc
Page 25 of 42
9 .
U S E F U L
I N F O R M A T I O N
A
– There is a
DEDUCTION (1)
key on the control panel that allows users to
shorten
the sterilizing
time (from 1 to 12 minutes) if required, while maintaining the length of the drying cycle
unchanged.
This key should be used once the program has been selected and immediately after the door has
been shut, when the flashing number indicating the sterilizing minutes appears on the display.
NOTE
The time reducing function should only be used when the chamber is loaded with
NON-
PACKAGED
instruments and articles!
B
– The machine has been designed to stop the sterilizing cycle in progress in the event of a power
failure. In this case, message
[rEt]
will appear on the display.
If this happens, it will be necessary to repeat the sterilizing process by opening the door and
activating the selected program again.
C
- Sterilizer mod.
STERILIX VACUUM PLUS, STERILIX STAR and STERILIX STAR FAST
has a bacteriological filter to preserve packaged instruments with sterile air.
D
– In compliance with EC standards, the autoclave has been fitted with a connection that allows a
PRINTER
to be installed (mod. M9002 without date stamp pr mod. M9003 with date stamp). This
allows users to print out certain parameters of fundamental importance (e.g. the progressive
number of the cycle). The printer produces a paper ticket certifying that the instruments have been
sterilized along with the relative date.