Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast
Rev. 09 of 18-04-02 - E:\Documenti1\Manuali superati non in uso\Vacuum Star GB-R09.doc
Page 26 of 42
1 0 .
K E Y
T O
D I S P L A Y E D
M E S S A G E S
01 - tPt
Indicates a block status.
02 - rEt
Indicates that there has been a power failure during the sterilizing phase.
03 - dor
The user is asked to open the door
04 - SLP
Displayed after 1 hour, when the door is open and the machine is not being used: turns
off the pre-heating element.
05 - Ptl
Indicates that the temperature is less than 110°C during the sterilizing phase, with a
pressure of more than 0.8 bar in the chamber: air in the chamber.
06 - btP
Displayed if 35 min. have elapsed and the temperature has not reached at least 115°C.
07 - PrS
Displayed during the sterilizing process: Pressure less than 0.8 bar with a temperature of
more than 120°C
08 - otr
Temperature in the boiler more than 142°C.
09 - Lot
Temperature in the boiler less than -5°C or probe broken.
10 - nPt
Temperature less than 50°C once 20 minutes have elapsed from the time the machine
was powered.
11 - Hit
Boiler temperature over 160° or probe wire broken, or faulty probe connection.
12 - dPr
The machine has not carried out the vacuum phase.
13 - Pr0
This message is displayed at the beginning of the cycle when no program has been
selected (select the program
without opening the door
).
14 - Pr1
Indicates that PROGRAM 1 has been selected. (134°)
15 - Pr2
Indicates that PROGRAM 2 has been selected. (121°)
16 - A-
Displayed during the DRYING PHASE:
A
= drying.
17 - Str
STERILIZED: indicates THAT THE STERILIZING OPERATIONS HAVE
TERMINATED AND THAT THE CYCLE CONFORMS. BEGINNING OF THE
UPKEEP CYCLE.
18 - 888
Temperature control test phase message.
19 - StP
Displayed during the PRINT phase (even when the printer is not connected).
20 - H2o
No water in fill tank (see filling procedure).
21 - FuL
Water level in tank sufficient for next cycle.
22 - PrE
Message lasting a few seconds. Precedes display of chamber temperature with door open
alone.
23 - tEP
( For technicians only ): Displays the temperature reached during the sterilizing phase.
24 - End
END OF STERILIZING CYCLE.
25 - PuP
( For technicians only ): Water fill pump only activated in test function. (7)
26 - rES
( For technicians only ): Heating element on to test real power draw of the heating
element. (13)
27 - Elb
( For technicians only ): N.O. solenoid. (14)
28 - Elc
( For technicians only ): Chamber water inlet solenoid valve. (7+16)