Packaging, Transport and Installation
15
4.4 Conditions for the Installation Site
Installation height: max. 2
000 m above sea level
Ambient temperature: 5 °C
– 40 °C
NOTICE
N7.0021
Ambient temperature
Temperatures outside the permitted range
Electronic and mechanical components may be damaged.
The performance data alter to an unknown extent.
•
Do not exceed or fall below the permitted temperature range (5 °C to
40 °C ambient temperature) of the device.
Maximum relative humidity < 80 % (at ambient temperatures
≤ 31 °C)
For ambient temperatures U
T
between 31 °C and 40 °C, the maximum relative humidity value L
F
linearly decreases according to L
F
=
–(U
T
– 55) / 0.3:
Ambient temperature
Max. rel. humidity
≤ 31 °C
80 %
33 °C
73.3 %
35 °C
66.7 %
37 °C
60 %
39 °C
53.3 %
40 °C
50 %
NOTICE
N8.0015
Humidity
High relative humidity
Electronic and mechanical components may be damaged.
The performance data alter to an unknown extent.
•
The relative humidity in the vicinity of the device should be kept as
low as possible.
4.5 Electrical Connection
WARNING
W2.0015
Danger to life through electric shock
Connection to sockets without protective earth
An electric shock can cause burns, cardiac arrhythmia, respiratory arrest,
as well as cardiac arrest.
•
The device may only be operated on sockets with protective earth
(PE).
Содержание AS 200 jet
Страница 2: ...2 Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany...
Страница 5: ...Notes on the Manual 5...
Страница 25: ...First Commissioning 25 Fig 10 Example of a label on the sieve lid...
Страница 42: ...Controlling the Device 42 Fig 21 Submenu SERVICE H7 1...
Страница 61: ...61...
Страница 62: ...62...
Страница 63: ...63...
Страница 64: ...Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany...