First Commissioning
20
5.2.1 Connecting the Manual Vacuum Regulation
Fig. 5:
Mounting the manual vacuum regulation
Insert the manual vacuum regulation (
G
) into the air outlet duct (
E
).
5.2.2 Adjustment of the Manual Vacuum Regulation
The manual vacuum regulation (
G
) has a lateral opening (
G2
) through which air is sucked in.
The size of the opening can be varied by means of a slider (
G1
). This allows to adjust the
desired negative pressure in the nozzle chamber (
A
).
When the opening is closed (
1
), the airflow from the nozzle (
B
) and thus also the negative
pressure in the nozzle chamber is at maximum. With the opening at maximum (
4
), the airflow
from the nozzle and thus the negative pressure in the nozzle chamber is at a minimum.
In between, the negative pressure can be infinitely adjusted by means of the slider (
G1
).
G
E