Controlling the Device
40
7.7.5 Time
Via the menu item "Time" the current time can be set. The device can be disconnected from the
mains for up to 30 days without losing the settings.
Use the rotary push button (
H4
) to navigate to the menu item "Time". The parameter is
highlighted in a darker background. The date is displayed in the format hour:minute:second
(hh:mm:ss) in three separate edit boxes.
Press the rotary push button to activate the first edit box (hour). The edit box is highlighted in
yellow.
Turn the rotary push button until the current hour is set in the 24-hour clock and confirm the
input by pressing the rotary push button. The next edit box is activated.
Set the current minute and confirm the input by pressing the rotary push button. The last edit
box is activated.
Set the current second. Press the rotary push button to save the input. The edit box is no
longer highlighted in yellow.
7.7.6 Signal Tone
The end of the sieving process, as well as occurring error messages can be signalled
acoustically by a warning tone.
Use the rotary push button (
H4
) to navigate to the menu item "Signal tone". The parameter is
highlighted in a darker background.
Press the rotary push button to activate the edit box. The edit box is highlighted in yellow.
Turn the rotary push button until the function is "On" or "Off" as desired.
Press the rotary push button to save the input. The edit box is no longer highlighted in
yellow.
The current setting is displayed by one of the following icons:
Signal tone activated
Signal tone deactivated
7.7.7 Interface
Under this menu item, the interface for the data transfer between the AS 200 jet and a PC can
be selected. It is possible to switch between the RS232 (
P
) and USB interface (
O
).
Use the rotary push button (
H4
) to navigate to the menu item "Interface". The parameter is
highlighted in a darker background.
Press the rotary push button to activate the edit box. The edit box is highlighted in yellow.
Turn the rotary push button until "RS232" or "USB" is displayed as desired.
Press the rotary push button to save the input. The edit box is no longer highlighted in
yellow.
The current setting is displayed by one of the following icons:
RS232 interface activated
USB interface activated
Содержание AS 200 jet
Страница 2: ...2 Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany...
Страница 5: ...Notes on the Manual 5...
Страница 25: ...First Commissioning 25 Fig 10 Example of a label on the sieve lid...
Страница 42: ...Controlling the Device 42 Fig 21 Submenu SERVICE H7 1...
Страница 61: ...61...
Страница 62: ...62...
Страница 63: ...63...
Страница 64: ...Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany...