Cleaning, Wear and Maintenance
48
Remove the feed hopper.
Unscrew the four M10x25 hex screws (
VS
) on the cover (
V
) at the back of the device.
Remove the cover (
V
) from the device.
Place the grease press on the lubrication points in turn (
W2, W3
) and press the right amount
of grease into the respective lubrication point.
Place the cover (
V
) back onto the device.
Secure the cover (
V
) using the four M10x25 hex screws (
VS
).
Assemble the feed hopper on the device.
7.3.2 Checking the limit switch
Fig. 32:
Checking the limit switch
NOTICE
The function of the limit switch (
L
) must be checked regularly every
6 months
.
Conducting the limit switch check
Switch the ABP 250 on (idling).
Open the lock (
M
) on the collecting receptacle (
N
).
Pull the collecting receptacle (
N
) out.
Result: the limit switch (
L
)
must
turn the drive motor (
O
)
off
.
Slide the collecting receptacle (
N
) back into the base frame.
Result: the drive motor (
O
)
does not start
. It can only be switched back on using the ON
switch.
Close the lock (
M
) on the collecting receptacle (
N
).
Switch the ABP 250 on using the ON switch.
M
N
O
L