4.  The horizontal scroll bar allows you to move the entire tracing itself along the X 

axis or to scroll along the pH tracing while viewing the pH values and time in the 
fields on the right hand side. To do the latter, the cursor must be engaged.  

 

 

 
 

5.  The cursors can be used to examine values at certain points, as well as to delete 

certain data from the analysis (Artefacts). 

 

 

 
 

The button in the center allows you to change the color and style of the cursor, the point 
style, the line style and width, display the name of the cursor, bring it to the center of the 
page, and go to it.  The lock button secures the specific cursor characteristics you have set.  
 
When trying to remove data, move the cursors to either side of the data set you want to 
remove. Ensure that the points are correct by looking at the values displayed. When you are 
certain that the points should be removed, click the button that says remove data between 
the cursors (the Artefact).  
 

 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание Dx-pH Measurement System

Страница 1: ...Dx pH Measurement System Model Number XXXXX Instructions for Use Respiratory Technology Corporation 13670 Danielson St Suite F Poway CA 9064 Ph 858 679 300 Fax 858 679 359...

Страница 2: ...2 Dx pH Measurement System User s Manual Version X X August 2005...

Страница 3: ...onditions are met 1 All repairs and modifications are carried out by Respiratory Technology Corporation or its assigned agents 2 Caution Changes or modifications not expressly approved by the party re...

Страница 4: ...udy Physician Setup Setup Calibrate the Probe Determining the Proper Catheter Placement Retrieving and Reviewing a Study Download pH Data Opening a Past Study Graphs and Reports Changing the pH Scale...

Страница 5: ...act with the receiver during the 24 48 hour pH study to record clinically relevant information such as meals and symptoms After completion of the pH study the pH data and patient information is downlo...

Страница 6: ...ng symbols are used on the Dx pH Probe and or in the accompanying documentation SYMBOL MEANING Lot Number Non Sterile Date of Manufacture Each device is for one 1 use only Expires Attention Read instr...

Страница 7: ...7 Dx Recorder Event Markers and Control Keys Event Marker Buttons Heartburn Button Cough Button Meals Button Supine Sleep Button Control Keys Up Arrow Key Down Arrow Key Escape Key...

Страница 8: ...be measures pH and the transmitter sends the data to the Dx Recorder Dx Recorder REPLACE PICTURE The Dx Recorder is a portable unit that receives pH data from the Dx pH Probe and Transmitter It saves...

Страница 9: ...PICTURE The Dx pH DataView Software displays pH data that is stored on the Data Card by the Dx Recorder Dx Calibration Vials and Dx Calibration Stand REPLACE PICTURE The Dx pH Probe is calibrated with...

Страница 10: ...by User Computer REPLACE PICTURE A PC with CD ROM Drive and Windows based operating system runs the Dx pH DataView Software program generates reports to be printed and stores patient data on the hard...

Страница 11: ...ent by the Dx Transmitter Upon completion of a study trained personnel will remove the data card from the recorder and insert it into a PC flash memory card reader The Dx Recorder is designed to be li...

Страница 12: ...o the card reader of your computer A window titled Removable Disk should appear close the window to continue In your Dx pH DataView program select New Patient from the File menu Enter the last name fi...

Страница 13: ...placement Setup Setup allows you to adjust the system time select the study duration and set user interface options You will be prompted to either accept or change time and date information If the tim...

Страница 14: ...Vials which are pre filled with pH buffer solution Select Cal from the main menu The recorder will display the Transmitter ID number that was used You will be prompted to accept this ID Pressing the u...

Страница 15: ...te Have the patient swallow sips of water to see if they can feel the probe in their throat if they can pull it back slightly until it is comfortable for the patient After the proper position has been...

Страница 16: ...for the card automatically opens simply close it and proceed with the following instructions In the File menu of the Dx pH DataView program select the option to Retrieve Completed Study Opening a Pas...

Страница 17: ...screen Changing the pH Scale The pH scale is set at a default of 0 10 To change this scale select Y Scale Unlock from the Y Scale Menu Now you can enter the desired values for the study in review Whe...

Страница 18: ...n the top right corner of the screen perform various functions to allow this manipulation a The button on the left allows you to move the cursors along the X axis When the cursor rests on a point it w...

Страница 19: ...o the left or the right 3 The button is for scrolling in fine increments to view pH changes with precision along the X or Y axis Note that to use it along the Y axis the Y Scale must be unlocked The t...

Страница 20: ...ts The button in the center allows you to change the color and style of the cursor the point style the line style and width display the name of the cursor bring it to the center of the page and go to...

Страница 21: ...21...

Страница 22: ...would like to perform the edit again click the button that says Undo Delete After deleting the specified data sets it will not be included in the analysis and will not be printed To replace this data...

Страница 23: ...23 You can right click on any point on the graph to display the following menu An event marker appears on the lower graph as follows...

Страница 24: ...ny of the three lines Meal Supine or Symptoms and right clicking your mouse will allow you to change the appearance of the lower graph The following menu will appear Fill Base Line will Selecting Comm...

Страница 25: ...e Width allows you to adjust the width of the graph lines Bar Plots The Interpolation function allows you to connect the points using various methods of interpolation illustrated in the six different...

Страница 26: ...ng doctor The patient name first and last and study date are required fields in specific formats The user defined events are displayed and if none were entered the words Event 1 Event 2 and Event 3 wi...

Страница 27: ...vious information will be removed from that card make sure the patient data has been saved Printing Graphs and Reports Selecting Print Graph from the Printing Menu prints the graph with no preview If...

Страница 28: ...28 Click OK when the desired settings have been selected...

Страница 29: ...ons attachment B How to keep a written diary For analysis information about episodes of pain and episodes of cough supine periods and meal periods is needed and is very important See attachment B How...

Страница 30: ...r evaluation Recorder will not accept the data card Data is still saved from previous study Download data into Dx ph DataView and reinitialize card Start pH Study Transmitter identified does not match...

Страница 31: ...he front cover of the Recorder Let the surfaces evaporate thoroughly before use Service The Dx Recorder does not require routine calibration or servicing of internal components unless the recorder bec...

Страница 32: ...32 Attachment A Patient Instructions...

Страница 33: ...r meals and sleep periods using the time from the clock displayed on the Dx Recorder This will ensure that the diary times entered in the Dx DataView software are correct when the patient s pH study i...

Страница 34: ...34...