P
AGE
6-2
T
RILOGY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
1002735, V
ER
. 05
•
Yellow Solid Indicator - When the device detects a low priority alarm, a solid yellow light appears
on the Alarm Indicator/Audio Pause button.
The Alarm Indicator/Audio Pause button does not light up when informational messages or confirmation alerts
display.
6.2.2 A
UDIBLE
I
NDICATORS
An audible indicator sounds whenever a power failure or a high, medium, or low priority alarm is detected.
Additionally, an audible indicator sounds for informational messages and to confirm that certain actions have
occurred (for example, when an SD card is inserted or removed from the device).
•
Ventilator Inoperative Audible Indicator
– When a ventilator inoperative alarm occurs, a contin-
uous audible alarm sounds. The alarm descriptions later in this chapter display this indicator
as:
•
Power Failure Audible Indicator
– When a power failure occurs, a series of beeps sounds in a 1
beep pattern, repeating one second on, then one second off. The alarm descriptions later in this
chapter display this indicator as:
• •
•
High Priority Audible Indicator
– When a high priority alarm is detected, a series of beeps sound
in the following pattern, which is repeated twice: 3 beeps, a pause, and then 2 more beeps. This
indicator continues until the cause of the alarm is corrected or the audible alarm is paused. The
alarm descriptions later in this chapter display this indicator as:
• • • • • • • • • •
•
Medium Priority Audible Indicator
– When a medium priority alarm is detected, a series of
beeps sound in a 3-beep pattern. This pattern repeats until the cause of the alarm is corrected or
the audible alarm is paused. The alarm descriptions later in this chapter display this indicator as:
• • •
•
Low Priority Audible Indicator
– When a low priority alarm is detected, a series of beeps sound
in a 2-beep pattern. This pattern repeats until the cause of the alarm is corrected or the audible
alarm is paused. The alarm descriptions later in this chapter display this indicator as:
• •
•
Informational Messages and Confirmation Audible Indicators
– When an informational mes-
sage appears on screen, a brief, 1- beep audible indicator sounds. Additionally, when the device
detects that a certain action has been completed (for example, when the Start/Stop button is
pressed to start therapy, or when an SD card is inserted or removed from the device) a brief, 1-
beep audible indicator sounds. The alarm descriptions later in this chapter display this indicator
as:
•
NOTE
For the alarm indicators noted throughout this manual, each “dot” represents an audible beep.
Содержание Trilogy 02
Страница 1: ...Trilogy 100 Trilogy 200 Trilogy 02 Trilogy 202 Service Technical Information ...
Страница 35: ...PAGE 3 20 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 This page intentionally left blank ...
Страница 49: ...PAGE 4 14 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 This page intentionally left blank ...
Страница 81: ...PAGE 5 32 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 This page intentionally left blank ...
Страница 161: ...PAGE 7 24 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 This page intentionally left blank ...
Страница 177: ...PAGE 8 16 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 3 4 5 ...
Страница 181: ...PAGE 8 20 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 6 ...
Страница 184: ...PAGE 8 23 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 210: ...PAGE 8 49 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 FIGURE 8 49 PROPER TUBING CONNECTIONS 2 ...
Страница 211: ...PAGE 8 50 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 FIGURE 8 50 PROPER TUBING CONNECTIONS 3 ...
Страница 212: ...PAGE 8 51 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 FIGURE 8 51 PROPER TUBING CONNECTIONS 4 ...
Страница 217: ...PAGE 8 56 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 FIGURE 8 57 PROPER TUBING CONNECTIONS 2 ...
Страница 218: ...PAGE 8 57 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 FIGURE 8 58 PROPER TUBING CONNECTIONS 3 ...
Страница 339: ...PAGE 10 36 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 2 Click on the Service Software link ...
Страница 343: ...PAGE 10 40 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 346: ...PAGE 10 43 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 351: ...PAGE 10 48 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 352: ...PAGE 10 49 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 353: ...PAGE 10 50 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 354: ...PAGE 10 51 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 377: ...PAGE 10 74 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 11 Click the System option 12 Click the Hardware tab ...
Страница 389: ...PAGE 10 86 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 3 Follow the steps on the screen shot below ...
Страница 399: ...PAGE 10 96 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 3 Follow the on screen steps shown below ...
Страница 403: ...PAGE 10 100 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 10 Follow the on screen prompts ...
Страница 409: ...PAGE 10 106 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 This page intentionally left blank ...
Страница 418: ...PAGE 12 5 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 419: ...PAGE 12 6 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 420: ...PAGE 12 7 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 421: ...PAGE 12 8 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 422: ...PAGE 12 9 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 423: ...PAGE 12 10 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 424: ...Reorder 1002735 Ver 05 BRM 10 11 2010 ...