P
AGE
3-8
T
RILOGY
S
ERVICE
& T
ECHNICAL
I
NFORMATION
1002735, V
ER
. 05
3.14.2 G
UIDANCE
AND
M
ANUFACTURER
’
S
D
ECLARATION
- E
LECTROMAGNETIC
I
MMUNITY
This device is intended for use in the electromagnetic environment specified below. The user of this device
should make sure it is used in such an environment.
IMMUNITY
TEST
IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT -
GUIDANCE
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast
Transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input-output
lines
±2 kV for supply mains
±1 kV for input/output
lines
Mains power quality should be
that of a typical home or
hospital environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common mode
±1 kV differential
mode
±2 kV for common
mode
Mains power quality should be
that of a typical home or
hospital environment.
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for
0.5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec
<5% U
T
(>95% dip in U
T
) for
0.5 cycle
40% U
T
(60% dip in U
T
) for 5
cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec
Mains power quality should be
that of a typical home or
hospital environment. If the user
of the device requires continued
operation during power mains
interruptions, it is
recommended that the device
be powered from an
uninterruptible power supply or
a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical home
or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Содержание Trilogy 02
Страница 1: ...Trilogy 100 Trilogy 200 Trilogy 02 Trilogy 202 Service Technical Information ...
Страница 35: ...PAGE 3 20 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 This page intentionally left blank ...
Страница 49: ...PAGE 4 14 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 This page intentionally left blank ...
Страница 81: ...PAGE 5 32 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 This page intentionally left blank ...
Страница 161: ...PAGE 7 24 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 This page intentionally left blank ...
Страница 177: ...PAGE 8 16 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 3 4 5 ...
Страница 181: ...PAGE 8 20 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 6 ...
Страница 184: ...PAGE 8 23 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 210: ...PAGE 8 49 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 FIGURE 8 49 PROPER TUBING CONNECTIONS 2 ...
Страница 211: ...PAGE 8 50 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 FIGURE 8 50 PROPER TUBING CONNECTIONS 3 ...
Страница 212: ...PAGE 8 51 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 FIGURE 8 51 PROPER TUBING CONNECTIONS 4 ...
Страница 217: ...PAGE 8 56 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 FIGURE 8 57 PROPER TUBING CONNECTIONS 2 ...
Страница 218: ...PAGE 8 57 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 FIGURE 8 58 PROPER TUBING CONNECTIONS 3 ...
Страница 339: ...PAGE 10 36 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 2 Click on the Service Software link ...
Страница 343: ...PAGE 10 40 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 346: ...PAGE 10 43 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 351: ...PAGE 10 48 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 352: ...PAGE 10 49 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 353: ...PAGE 10 50 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 354: ...PAGE 10 51 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 377: ...PAGE 10 74 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 11 Click the System option 12 Click the Hardware tab ...
Страница 389: ...PAGE 10 86 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 3 Follow the steps on the screen shot below ...
Страница 399: ...PAGE 10 96 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 3 Follow the on screen steps shown below ...
Страница 403: ...PAGE 10 100 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 10 Follow the on screen prompts ...
Страница 409: ...PAGE 10 106 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 This page intentionally left blank ...
Страница 418: ...PAGE 12 5 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 419: ...PAGE 12 6 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 420: ...PAGE 12 7 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 421: ...PAGE 12 8 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 422: ...PAGE 12 9 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 423: ...PAGE 12 10 TRILOGY SERVICE TECHNICAL INFORMATION 1002735 VER 05 ...
Страница 424: ...Reorder 1002735 Ver 05 BRM 10 11 2010 ...