User Manual
39
A
PPENDIX
A: EMC I
NFORMATION
G
UIDANCE
AND
M
ANUFACTURER
’
S
D
ECLARATION
- E
LECTROMAGNETIC
E
MISSIONS
:
Th
is device is intended for use in the
electromagnetic environment specifi ed below. Th
e user of this device should make sure it is used in such an environment.
E
MISSIONS
T
EST
C
OMPLIANCE
E
LECTROMAGNETIC
E
NVIRONMENT
- G
UIDANCE
RF emissions
CISPR 11
Group 1
The device uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The device is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fl uctuations/Flicker emissions
IEC 61000-3-3
Complies
G
UIDANCE
AND
M
ANUFACTURER
’
S
D
ECLARATION
- E
LECTROMAGNETIC
I
MMUNITY
:
Th
is device is intended for use in the
electromagnetic environment specifi ed below. Th
e user of this device should make sure it is used in such an environment.
I
MMUNITY
T
EST
IEC 60601 T
EST
L
EVEL
C
OMPLIANCE
L
EVEL
E
LECTROMAGNETIC
E
NVIRONMENT
-
G
UIDANCE
Electrostatic Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic
tile. If fl oors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical Fast Transient/Burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input-output lines
±2 kV for supply mains
±1 kV for input/output lines
Mains power quality should be that of a
typical home or hospital environment.
Surge
IEC 61000-4-5
±1 kV diff erential mode
±2 kV common mode
±1 kV diff erential mode
±2 kV for common mode
Mains power quality should be that of a
typical home or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
) for 5 cycles
70% U
T
(30% dip in U
T
) for 25 cycles
<5% U
T
(>95% dip in U
T
) for 5 sec
<5% U
T
(>95% dip in U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
) for 5 cycles
70% U
T
(30% dip in U
T
) for 25 cycles
<5% U
T
(>95% dip in U
T
) for 5 sec
Mains power quality should be that of a
typical home or hospital environment. If
the user of the device requires continued
operation during power mains interruptions,
it is recommended that the device be
powered from an uninterruptible power
supply or a battery.
Power frequency (50/60 Hz)
magnetic fi eld
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fi elds should be
at levels characteristic of a typical location in
a typical hospital or home environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.