38
User Manual
P
RESSURE
Output:
4 to 30 cm H
2
O
C
ONTROL
A
CCURACY
Parameter
Range
Accuracy
Max Pressure
4 to 30 cm
H
2
O
*
Min EPAP
4 to 25 cm
H
2
O
*
Max EPAP
4 to 25 cm
H
2
O
*
Min Pressure Support
0 to 26 cm
H
2
O
*
Max Pressure Support
0 to 26 cm
H
2
O
*
Breath Rate
4 to 30 BPM
Greater of ± 1 BPM or
± 10% of the setting (when
measured over a 4 minute
period)
Timed Inspiration
0.5 to 3.0 seconds
± (0.1 + 10% of the setting)
seconds
Ramp Duration
0 to 45 minutes
± 10% of the setting
Rise Time
1 to 6 **
± 25%
***
*
Dynamic pressure accuracy is ± 1.5 cm H
2
O measured at the patient end of the
circuit with a Whisper Swivel II and varying fl ow conditions. Static pressure accuracy
is ± 1.25 cm H
2
O measured at the patient end of the circuit with a Whisper Swivel II
and no patient fl ow.
** Th
e range of values correspond to tenths of seconds (e.g., a setting of 4
indicates a Rise Time of 0.4 seconds).
*** Measured at the patient end of circuit with a Whisper Swivel II exhalation
device and no patient fl ow.
M
EASURED
P
ARAMETER
A
CCURACY
Parameter
Accuracy
Respiratory Rate
Greater of ±1 BPM or ±10% of reading when
measured over a four minute period
Exhaled Tidal Volume
± (25 + 0.15 of reading) ml
Exhaled Minute
Ventilation
± (1 + 0.15 of reading) L/min
Leak Rate
± (5 + 0.15 of reading) L/min
WEEE/R
O
HS R
ECYCLING
D
IRECTIVES
If you are subject to the WEEE/RoHS recycling directives, refer to www.respironics.com for the
passport for recycling this product.