38
IX. Service
CAUTION
•
920 Hand Held Pulse Oximeters are
sensitive and must be repaired by
knowledgeable and specially trained
personnel only. Any sign or evidence of
opening the system, field service by non-
Respironics personnel, tampering, or any
kind of misuse or abuse of the system,
shall void the warranty in its entirety.
The advanced digital circuitry within the 920 requires no periodic
maintenance or calibration other than battery replacement.
Respironics does not recommend field repair of the 920 Hand
Held Pulse Oximeter.
The circuit board in the 920 series is a multi-
layer board using traces 0.01” wide. Due to the very small trace
size, extreme care must be used when replacing components to
prevent permanent non-repairable damage to the circuit board.
Most components are surface-mounted and require special hot air
jet soldering and desoldering equipment. After any repairs are
made, the pulse oximeter must be tested to ensure correct
operation.
NOTE:
All repair work on the 920 should be done by trained
Respironics personnel. For Respironics Technical
Support contact:
Technical Support Telephone: 1-800-421-8754
(USA and Canada)
Technical Support Telephone: 1-770-499-1212
(International)
Customer Service Telephone: 1-800-345-6443
(USA and Canada)
Customer Service Fax: 1-800-886-0245
(USA and Canada)
www.respironics.com
All non-warranty work shall be done according to Respironics
standard rates and charges in effect at the time of delivery to
Respironics. All repairs include an 18-hour burn-in and a complete
retest of the pulse oximeter using factory test fixtures.