G. Cleaning the Sensors
CAUTION
•
Do not immerse the sensors in liquid
to clean.
•
Do not use caustic or abrasive
cleaning agents.
Carefully clean the Respironics reusable sensors with an isopropyl
alcohol wipe and ensure that all tape residue is removed. Allow enough
time for the sensor to dry thoroughly before reusing. The sensors may
be sterilized using ethylene oxide (EtO) (cold cycle).
H. Sensor Compatibility
WARNING
•
Use only Respironics-manufactured
sensors. These sensors are
manufactured to meet the calibration
requirements for Respironics Pulse
Oximeters.
The 920 is compatible with all Respironics-manufactured sensors
(except the Fiber Optic sensors). Refer to Tables V, VI, and VII for a
list of the sensors, available sensor accessories, and their model
numbers.