3. Setup Mode
Setup mode is used to control the internal time-of-day clock and the
external real-time printer (purchased separately). The setup mode is
initiated by holding the “
ñ
” switch when the unit is turned on by
pressing the “
” switch. In setup mode, the “
” switch and the “
ñ
”
switch are used to make the selections.
Advance to the next sequential mode by pressing the “
” switch.
Each time the “
ñ
” switch is pressed, the number on the
♥
display
will increment. It starts with the current value stored in memory for
the parameter designated in the SpO
2
display. When the correct
value appears in the
♥
display, pressing the “
” switch will advance
the SpO
2
display to the next sequential parameter as listed in Table
I. This process is continued until all parameters are set. The
settings can be easily checked, since the first value displayed for
each parameter represents the current setting. When the setting
sequence has been completed, the 920 exits the setup mode and
begins normal operation.
Range of Values
Sequence
Appears in
SpO
2
Display
From:
To:
Printer
“
Prn
”
00
15
Year
“
y
”
00
99
Month
“
nn
”
00
12
Day
“
d
”
01
31
Hours
“
h
”
00
23
Minutes
“
nn
”
00
59
Table I: Printer, clock, and calendar mode parameters.
a. Printer Settings
NOTE:
The 920P prints out real-time data only. Data
stored in the 920M can not be downloaded to
the 920P.
“
Prn
” will appear in the SpO
2
display indicating print setup mode.
There are 16 options for the printer mode: 00 through 15. Each printer
mode is explained in Table II. The modes determine how often data
is written out to the printer and the format of data written to the
printer. Refer to Figure V for a flowchart of setting the printer mode.
When the printer setting sequence has been completed, the 920
continues to the calendar settings (refer to section b of this section).