English
15
For more information on the SD card refer to the SD card protective folder provided with your
device.
Note:
The SD card should not be used for any other purpose.
Travelling
You can take your Lumis HFT device with you wherever you go. Just keep the following in mind:
•
Use the travel bag provided to prevent damage to the device.
•
Empty the humidifier before packing it in the travel bag.
•
Make sure you have the appropriate power cord for the region you are travelling to. For
information on purchasing, contact your care provider.
CAUTION
Do not use the device on a plane or in a car due to the risk of inhalation of water.
Troubleshooting
If you have any problems, have a look at the following troubleshooting topics. If you are not able to
fix the problem, contact your care provider or ResMed. Do not try to open the device.
General troubleshooting
Problem/possible cause
Solution
I am getting a dry or blocked nose
Humidity level may be set too low.
Adjust the Humidity Level.
I am getting droplets of water in the nasal cannula and air tubing
Humidity level may be set too high.
Adjust the Humidity Level.
Tubing wrap not in use.
Fit tubing wrap.
See the tubing wrap user guide for fitting instructions.
Air flow in my nose seems too high (it feels like I am getting too much air)
Flow may be set too high.
Talk to your care provider about your settings.
Air flow in my nose seems too low (it feels like I am not getting enough air)
Flow may be set too low.
Talk to your care provider about your settings.
My screen is black
Backlight on the screen may have turned off. It turns off
automatically after a short period of time.
Press Home or the dial to turn it back on.
Power may not be connected.
Connect the power supply and make sure the plug is fully
inserted.
I have stopped therapy, but the device is still blowing air
Device is drying circuit and nasal cannula.
Device blows a small amount of air in order to dry the air
tubing and nasal cannula. It will stop automatically after 90
minutes.