Alarms and troubleshooting
NOT013281-8
Relative alarm priorities
6.4 Alarm settings
CAUTION
If you change any of the alarm settings, see Section 6.5 on page 74 for instructions on
testing the alarms.
: Non-invasive ventilation;
: Invasive ventilation
High-priority alarms
Medium-priority alarms
Low-priority alarms
Highest priority
Lowest priority
LP – Low Flow?
EXT BATT EMPTY
NiMH battery
LP – CIRCUIT OPEN
INT BATT LOW
EXT BATT MISSING
MAXI P inspi
LOW PEEP
EXT BATT LOW
O
2
FAULT
PATIENT APNEA
EXT MISSING
MAXI LEAKS
MAINS MISSING
MAXI P expi
EXPI resist
WRONG CIRCUIT
INSPI resist
MINI V/MIN
MAXI V/MIN
MINI Vt
MAXI F
MAXI Vt
ADJUST CLOCK
FIO
2
21
VENTIL STOPPED
MAXI FIO
2
REPLACE BLOWER
MINI FIO
2
REPLACE
O
2
CELL
PRESENCE O
2
HP
INT BATT EMPTY
TEST FAILED
+
-
Default value
Settings range
Value:
Minimum
Maximum
Increment
Patient type:
Child
Adult
Child
Adult
Child
Adult
Max. resp. rate
(bpm)
35
15
10
99
1
Max. pressure
(cm H
2
O)
30
40
5
99
1
Maxi e
(L/min)
5
15
3.5
19.5 / NO
39.5 / NO
0.5
Mini e
(L/min)
1.0
3
0.2
10.0
20.0
0.2
Maxi i
(L/min)
5.0
15
3.5
19.5 / NO
39.5 / NO
0.5
Mini i
(L/min)
1.0
3
0.2
10.0
20.0
0.2
Maxi Vte (mL)
750
550
1000
50
200
10
200
600
750 / NO
1000
50
1000
2400 / NO
100
°
V
°
V
°
V
°
V
Содержание Elisee 350
Страница 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Страница 2: ...Making quality of care easy Respiratory Care solutions...
Страница 4: ......
Страница 6: ...iv...
Страница 10: ...4 102 NOT013281 8...
Страница 20: ...14 102 NOT013281 8...
Страница 50: ...44 102 NOT013281 8...
Страница 86: ...80 102 NOT013281 8...
Страница 109: ...Making quality of care easy Respiratory Care solutions...