Ventilation
NOT013281-8
Triggers used with a double circuit
: Non-invasive ventilation;
: Invasive ventilation
Note:
For the non-invasive inspiratory trigger:
•
Trigger = 1 is the maximum sensitivity value (triggering of a cycle with minimal patient
effort) and Trigger = 5 is the minimum sensitivity value (the patient must make a greater
effort to trigger a cycle).
•
Trigger = Auto corresponds to a value of 3.
For the expiratory trigger, the Auto value also corresponds to a value of 3.
Triggers used with a single circuit
: Non-invasive ventilation;
: Invasive ventilation
Default values
Minimum
Maximum
Increment
Paediatric
Adult
Paediatric
Adult
Paediatric
Adult
Inspiratory
pressure
trigger
(cm H
2
O)
ACV
PACV
NO
NO
0.2
0.2
5.9/NO
5.9/NO
0.1
PS
PS.Vt
SIMV
PSIMV
NO
NO
0.2
0.2
6
6
0.1
Inspiratory
flow trigger
(L/min)
ACV
PACV
1.5
3
0.2
0.2
9.9/NO
9.9/NO
0.1
PS
PS.Vt
SIMV
PSIMV
1.5
3
0.2
0.2
10.0
10.0
0.1
Non-invasive
inspiratory
trigger
ACV
PACV
3
3
1
1
5/NO
5/NO
1
PS
PS.Vt
SIMV
PSIMV
3
3
1
1
5
5
1
Expiratory
trigger
PS
PS.Vt
SIMV
PSIMV
AUTO
25
AUTO
25
10
10
90/AUTO
90/AUTO
1
Default values
Minimum
Maximum
Increment
Paediatric
Adult
Paediatric
Adult
Paediatric
Adult
Inspiratory
pressure
trigger
(cm H
2
O)
ACV
PACV
0.5
1.0
0.2
0.2
5.9/NO
5.9/NO
0.1
PS
PS.Vt
SIMV
PSIMV
0.5
1.0
0.2
0.2
6
6
0.1
Non-invasive
inspiratory
trigger
ACV
PACV
3
3
1
1
5/NO
5/NO
1
PS
PS.Vt
SIMV
PSIMV
3
3
1
1
5
5
1
Expiratory
trigger
PS
PS.Vt
SIMV
PSIMV
AUTO
25
AUTO
25
10
10
90/AUTO
90/AUTO
1
Содержание Elisee 350
Страница 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Страница 2: ...Making quality of care easy Respiratory Care solutions...
Страница 4: ......
Страница 6: ...iv...
Страница 10: ...4 102 NOT013281 8...
Страница 20: ...14 102 NOT013281 8...
Страница 50: ...44 102 NOT013281 8...
Страница 86: ...80 102 NOT013281 8...
Страница 109: ...Making quality of care easy Respiratory Care solutions...