Maintenance and disinfection
NOT013281-8
7 Maintenance and disinfection
WARNING
The device should be cleaned and disinfected before it is used for the first time, and also
between patients.
The disinfection of the ventilator must be carried out as stated in the manufacturer’s instructions
and in accordance with the current protocol of the health care institution.
Service personnel must be familiar with cleaning and disinfection protocols as well as the
precautions required for certain materials.
The following figure shows the components which need to be cleaned and/or replaced:
7.1 Maintenance schedule
Any recommendations concerning cleaning frequency and the replacement of disposable
components are provided by the manufacturers of those components. As an indication, this table
provides information on the components that must be cleaned before reuse.
+
Antibacterial filter
Expiratory valve for double
circuit
Clapper
Dust filter
Oxygen
sensor
Ventilator casing
Patient circuit
Maintenance frequency
For the same patient
Between patients
Patient circuit
Hospital protocol or manufacturer
Yes
Antibacterial filter
If used for antiviral and antibacterial
protection, refer to the manufacturer’s
instructions
Yes
Expiratory valve for double circuit
Hospital protocol or one week
•
With antibacterial filter: hospital
protocol or one week
•
Without filter: clean between patients
Dust filter
Every 6 months at least
Nebuliser
Hospital protocol or manufacturer
Yes
Oxygen sensor
Refer to ventilator alarm if necessary
Device casing
Hospital protocol or one week
Yes
Содержание Elisee 350
Страница 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Страница 2: ...Making quality of care easy Respiratory Care solutions...
Страница 4: ......
Страница 6: ...iv...
Страница 10: ...4 102 NOT013281 8...
Страница 20: ...14 102 NOT013281 8...
Страница 50: ...44 102 NOT013281 8...
Страница 86: ...80 102 NOT013281 8...
Страница 109: ...Making quality of care easy Respiratory Care solutions...