Alarms and troubleshooting
NOT013281-8
* In pressure-controlled modes, the Pmax alarm threshold cannot be set to a value lower than the delivered Pi + PEEP (or the
delivered pressure support). If the user sets an alarm threshold lower than this pressure value, the alarm threshold setting will
be ignored. In this situation, the delivered Pi + PEEP (or PS + PEEP) becomes the Pmax alarm threshold.
Note:
The Battery Flat and Battery Low alarms are deactivated as soon as the Elisée 350 is
connected to the mains or to an external DC power supply. Press
to remove the alarm
message from the screen.
MINI FIO
2
Minimum FiO
2
High
FiO
2
– 10%
[NO]
Maximum of 20 consecutive
cycles below alarm threshold
MAXI FIO
2
Maximum FiO
2
High
FiO
2
+ 10%
[NO]
Maximum of 20 consecutive
cycles above alarm threshold
O
2
FAULT
HP O
2
missing
High
FiO
2
≠ 21% or
button pressed and
HP O
2
≥ 0.5 bar
[2 min]
Immediately
WRONG
CIRCUIT
Device configured for single
circuit (manual test), but a
double circuit is connected
High
V
TE
> 10% of V
TI
[2 min]
6 cycles
MAXI LEAKS
High Leak
High
Alarm setting
[2 min]
6 consecutive cycles above
alarm threshold
PATIENT
APNEA
Apnea
Medium
Tapnea
[2 min]
Immediately
INT BATT
EMPTY
Internal Battery Flat
High
Charge < 5%
[NO]
Immediately
INT BATT
LOW
Internal Battery Low
Medium
Charge < 15%
[2 min]
Immediately
EXT BATT
EMPTY
External Battery Flat
Medium
Charge < 5%
[2 min]
Immediately
EXT BATT
LOW
External Battery Low
Low
Charge < 15%
[2 min]
Immediately
EXT BATT
MISSING
No External Battery
Low
-
[YES]
Immediately
EXT MISSING
External Power Supply
Disconnection
Low
-
[YES]
1 s
MAINS
MISSING
Mains Disconnection
Low
-
[YES]
1 s
TEST FAILED
Manual Test Interrupted
High
If one of the four
test steps fails
[2 min]
-
Symbol
Message
Alarm
Priority
Trigger
threshold [can
be stopped?]
Time to trigger
O
2
100%
2 min
Содержание Elisee 350
Страница 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Страница 2: ...Making quality of care easy Respiratory Care solutions...
Страница 4: ......
Страница 6: ...iv...
Страница 10: ...4 102 NOT013281 8...
Страница 20: ...14 102 NOT013281 8...
Страница 50: ...44 102 NOT013281 8...
Страница 86: ...80 102 NOT013281 8...
Страница 109: ...Making quality of care easy Respiratory Care solutions...