English 17
Air filter
Standard:
Material: Polyester non woven fiber
Average arrestance: >75% for ~7 micron dust
Hypoallergenic:
Material: Acrylic and polypropylene fibers in a polypropylene
carrier
Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron
dust
Aircraft use
ResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21,
category M) for all phases of air travel.
Wireless module
Technology used:
2G GSM
It is recommended that the device is a minimum distance of 2 cm from the body during operation. Not applicable to masks,
tubes or accessories
Operating pressure range
AutoSet, CPAP:
4 to 20 cm H
2
O (4 to 20 hPa)
Supplemental oxygen
Maximum flow:
4 L/min
Pneumatic flow path
1. Flow sensor
2. Blower
3. Pressure sensor
4. Mask
5. Air tubing
6. Humidifier
7. Device
8. Inlet filter
Design life
Device, power supply unit:
5 years
Cleanable humidifier:
2.5 years
Air tubing:
6 months
General
The patient is an intended operator.
Humidifier performance
The following settings have been tested at 22°C ambient temperature:
Mask Pressure
cm H
2
O (hPa)
RH output %
Nominal system output AH
1
, BTPS
2
Setting 4
Setting 8
Setting 4
Setting 8
4 85 100
6 >10
10 85 100 6 >10
20 85 90 6 >10
1
AH - Absolute Humidity in mg/L
2
BTPS - Body Temperature Pressure Saturated