16
Environmental conditions
Operating temperature:
+5°C to +35°C
Note:
The air flow for breathing produced by this therapy
device can be higher than the temperature of the room.
Under extreme ambient temperature conditions (40°C) the
device remains safe.
Operating humidity:
10 to 95% relative humidity, non-condensing
Operating altitude:
Sea level to 2,591 m; air pressure range 1013 hPa to
738 hPa
Storage and transport temperature:
-20°C to +60°C
Storage and transport humidity:
5 to 95% relative humidity, non-condensing
Electromagnetic compatibility
The AirSense 10 complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-
2:2007, for residential, commercial and light industry environments. It is recommended that mobile communication devices
are kept at least 1 m away from the device.
Information regarding the electromagnetic emissions and immunity of this ResMed device can be found on
www.resmed.com, on the Products page under Service and Support.
EN 60601-1:2006 classification
Class II (double insulation), Type BF, Ingress protection IP22.
Sensors
Pressure sensor:
Internally located at device outlet, analog gauge pressure
type, -5 to +45 cm H
2
O (-5 to +45 hPa)
Flow sensor:
Internally located at device inlet, digital mass flow type, -70
to +180 L/min
Maximum single fault steady pressure
Device will shut down in the presence of a single fault if the steady state pressure exceeds:
30 cm H
2
O (30 hPa) for more than 6 sec or 40 cm H
2
O (40 hPa) for more than 1 sec.
Sound
Pressure level measured according to EN ISO 17510-1:2009 (CPAP mode):
SlimLine:
26.6 dBA with uncertainty of 2 dBA
Standard:
26.6 dBA with uncertainty of 2 dBA
Power level measured according to EN ISO 17510-1:2009 (CPAP mode):
SlimLine:
34.6 dBA with uncertainty of 2 dBA
Standard:
34.6 dBA with uncertainty of 2 dBA
Declared dual-number noise emission values in accordance with ISO 4871:1996.
Physical - device and humidifier
Dimensions (H x W x D):
116 mm x 255 mm x 150 mm
Air outlet (complies with ISO 5356-1:2004):
22 mm
Weight (device and cleanable humidifier):
1248 g
Housing construction:
Flame retardant engineering thermoplastic
Water capacity:
To maximum fill line 380 mL
Cleanable humidifier - material:
Injection molded plastic, stainless steel and silicone seal
Temperature
Maximum heater plate:
68°C
Cut-out: 74°C
Maximum gas temperature:
41°C