English 13
Problem/possible cause
Solution
Data transfer is not enabled for your device.
Contact your care provider to enable the data transfer
service.
My screen and buttons are flashing
Software upgrade is in progress.
Software upgrade takes approximately 10 minutes to
complete.
Device messages
Device message/possible cause
Solution
High leak detected, connect your water tub, tub seal or side cover
Humidifier may not be inserted properly.
Make sure the humidifier is correctly inserted.
Humidifier seal may not be inserted properly.
Open the humidifier and make sure that the seal is correctly
inserted.
High leak detected, connect your tubing
Air tubing may not be connected properly.
Make sure the air tubing is firmly connected at both ends.
Mask may be fitted incorrectly.
Make sure your mask is fitted correctly. See your mask user
guide for fitting instructions or use the Mask Fit function to
check your mask fit and seal.
Tubing blocked, check your tubing
Air tubing may be blocked.
Check the air tubing and remove any blockages. Press the
dial to clear the message and then press Start/Stop to
restart the device.
SD card error, remove your card and press Start to begin therapy
SD card may not be inserted correctly.
Remove and reinsert the SD card.
Read only card, please remove, unlock and re-insert SD card
SD card switch may be in the lock (read-only) position.
Move the switch on the SD Card from the lock position
to
the unlock position
and then re-insert it.
System fault, refer to user guide, Error 004
Device may have been left in a hot environment.
Allow to cool before re-use. Disconnect the power supply
and then reconnect it to restart the device.
Air filter may be blocked.
Check the air filter and replace it if there are any blockages.
Disconnect the power supply and then reconnect it to restart
the device.
Air tubing may be blocked.
Check the air tubing and remove any blockages. Press the
dial to clear the message and then press Start/Stop to
restart the device.
There may be water in the air tubing.
Empty the water from the air tubing. Disconnect the power
supply and then reconnect it to restart the device.