background image

10

 

 

Therapy data 

Your AirSense 10 device records your therapy data for you and your care provider so they can view 
and make changes to your therapy if required. The data is recorded and then transferred to your care 
provider wirelessly or via an SD card. 

 

Data transmission 

Your AirSense 10 device has the capability of wireless communication so that your therapy data can 
be transmitted to your care provider to improve the quality of your treatment. This is an optional 
feature that will only be available if you choose to benefit from it. It also allows your care provider to 
update your therapy settings in a more timely manner or upgrade your device software to ensure 
you receive the best therapy possible. 

The data is usually transmitted after therapy has stopped. In order to make sure that your data is 
transferred, leave your device connected to the mains power at all times and make sure that it is not 
in Airplane Mode. 

 

Notes:

 

 

Therapy data might not be transmitted if you use it outside of the country or region of purchase. 

 

Devices with wireless communication might not be available in all regions. 

 

SD card 

An alternative way for your therapy data to be transferred to your care provider is via the SD card. 
Your care provider may ask you to send the SD card by mail or to bring it in. When instructed by 
your care provider, remove the SD card. 

 

Do not remove the SD card from the device when the SD light is flashing. 

 

To remove the SD card: 

 

 
 

1.

 

Open the SD card cover. 

2.

 

Push in the SD card to release it. Remove the SD card from the device. 
Place the SD card in the protective folder and send it back to your care provider. 

 

For more information on the SD card refer to the SD card protective folder provided with your 
device. 

 

Note:

 The SD card should not be used for any other purpose. 

 

Traveling 

You can take your AirSense 10 device with you wherever you go. Just keep the following in mind: 

 

 

Use the travel bag provided to prevent damage to the device. 

 

 

Empty the humidifier and pack it separately in the travel bag. 

 

Содержание AirSence 10

Страница 1: ...User guide English User guide...

Страница 2: ...e AirSense 10 Elite device is indicated for the treatment of obstructive sleep apnea OSA in patients weighing more than 30 kg It is intended for home and hospital use The humidifier is intended for si...

Страница 3: ...es available for use with the device including Air tubing heated and non heated ClimateLineAir SlimLine Standard Filter Hypoallergenic filter standard filter Air10 DC DC converter SD card reader Air10...

Страница 4: ...o select an option Turn to adjust a selected option and press to save your change Home button Press to return to the Home screen Different icons may be displayed on the screen at different times inclu...

Страница 5: ...end into the power outlet 3 Connect the air tubing firmly to the air outlet located on the rear of the device 4 Open the humidifier and fill it with water up to the maximum water level mark Do not fil...

Страница 6: ...restored The AirSense 10 device has a light sensor that adjusts the screen brightness based on the light in the room Stopping therapy 1 Remove your mask 2 Press Start Stop or if SmartStart is enabled...

Страница 7: ...nning of therapy more comfortable Ramp Time is the period during which the pressure increases from a low start pressure to the prescribed treatment pressure You can set your Ramp Time to Off 5 to 45 m...

Страница 8: ...rts blowing air 3 Adjust the mask mask cushion and headgear until you get a Good result To stop Mask Fit press the dial or Start Stop If you are unable to get a good mask seal talk to your care provid...

Страница 9: ...sk then gently pull apart Cleaning You should clean the device weekly as described Refer to the mask user guide for detailed instructions on cleaning your mask 1 Wash the humidifier and air tubing in...

Страница 10: ...lockages by dirt or dust To replace the air filter 1 Open the air filter cover and remove the old air filter The air filter is not washable or reusable 2 Place a new air filter onto the air filter cov...

Страница 11: ...s not in Airplane Mode Notes Therapy data might not be transmitted if you use it outside of the country or region of purchase Devices with wireless communication might not be available in all regions...

Страница 12: ...rplane Mode To turn on Airplane Mode 1 In My Options turn the dial to highlight Airplane Mode and then press the dial 2 Turn the dial to select On and then press the dial to save the change The Airpla...

Страница 13: ...ed I have stopped therapy but the device is still blowing air Device is cooling down Device blows a small amount of air in order to avoid condensation in the air tubing It will stop automatically afte...

Страница 14: ...Air tubing may be blocked Check the air tubing and remove any blockages Press the dial to clear the message and then press Start Stop to restart the device SD card error remove your card and press St...

Страница 15: ...d into the pivot hole of the base 2 Slide the other side down the ridge until it clicks into place General warnings and cautions WARNING Make sure that you arrange the air tubing so that it will not t...

Страница 16: ...omeone is likely to trip over the power cord Blocking the air tubing and or air inlet of the device while in operation could lead to overheating of the device Keep the area around the device dry clean...

Страница 17: ...otection IP22 Sensors Pressure sensor Internally located at device outlet analog gauge pressure type 5 to 45 cm H2O 5 to 45 hPa Flow sensor Internally located at device inlet digital mass flow type 70...

Страница 18: ...operation Not applicable to masks tubes or accessories Operating pressure range AutoSet CPAP 4 to 20 cm H2O 4 to 20 hPa Supplemental oxygen Maximum flow 4 L min Pneumatic flow path 1 Flow sensor 2 Blo...

Страница 19: ...o 150 L min positive flow Leak2 12 L min or 20 of reading whichever is greater 0 to 60 L min 1 Results are expressed at ATPD Ambient Temperature and Pressure Dry 2 Accuracy may be reduced by the prese...

Страница 20: ...reduce pressure on natural resources and prevent hazardous substances from damaging the environment If you need information on these disposal systems please contact your local waste administration The...

Страница 21: ...eing spilled on or into an electronic device Warranty is void on product sold or resold outside the region of original purchase Warranty claims on defective product must be made by the initial consume...

Страница 22: ...ee Ave Milton Park Abingdon Oxfordshire OX14 4RW UK See www resmed com for other ResMed locations worldwide For patent information see www resmed com ip Air10 AirSense AutoSet ClimateLine HumidAir and...

Отзывы: