Integra 3™ Care and Maintenance
67
Research Instruments Ltd
11
Thermosafe Air Heating System
The Thermosafe Air Heating System contains a filter which can be easily
replaced by the user. The filter is designed to prevent foreign objects and
debris entering the Thermosafe Air Heating System, which could cause
a degradation in performance. In all cases the Integra 3 manipulation
system is expected to be used in a clean environment. Frequency of filter
changes will be determined by the level of cleanliness. However, as a
starting point, it is recommended that the filter should be replaced every
6 months.
Switch the Integra 3 off before replacing the filter.
Ensure that a filter is fitted at all times and that the filter housing
is correctly fitted.
Figure 11-2
Pull cover
to remove
Filter