Section 9
Syringes
48
Research Instruments Ltd
9
Figure 9-14
Figure 9-15
Remove Cap
Plunger Tip
9. Now press the syringe plunger slowly to fill the cylinder and the
tubing. Next fill the port and replace the cap (see Figure 9-11 and
Figure 9-12 on page 47).
7. Stand the SOS on end as shown below. Turn the Control knob to
move the plunger below the fill port. Remove the cap from the fill
port. Fill the syringe with oil ensuring that there are no air bubbles
in the syringe.
8. Add a few drops of oil to the fill port and insert the syringe
ensuring that no air bubbles are trapped. If an air bubble is
seen then reattach the syringe or expel the bubble into the
cylinder and allow it to rise to the top before continuing.