Section 10
Integra 3™ Troubleshooting
50
Research Instruments Ltd
10
SECTION 10 - TROUBLESHOOTING
For any issues not covered below, please contact your distributor or the
RI service team directly (see page 71 for RI contact details).
Problem
Possible cause
Solution
The stiffness of the Fine
Control Lever is too tight
or too loose
Lever stiffness not
adjusted
Select the green-ended hexagon driver from
the tool kit provided. Loosen the locking
screw, then adjust the lever’s resistance by
turning anti-clockwise to loosen or clockwise
to tighten.* (See Figure 10-1)
The stiffness of the
Coarse Control Lever is
too tight or too loose
Lever stiffness not
adjusted
Select the green-ended hexagon driver from
the tool kit provided. Insert the driver into
a screw head located around the Coarse
Control Lever and slightly tighten or loosen
the screw (less than one turn). Test the lever
action. If it is not to your liking, insert the
driver into a different screw head and loosen
or tighten. ** (See Figure 10-2)
Manipulation
* When adjusting the resistance, turn the adjuster by the smallest possible amount. If any problems occur
contact the RI service team.
** The screw must not be too loose or lost motion will be introduced.
Tighten/Loosen
these three
screws.
Figure 10-1
Figure 10-2
Locking Screw
Adjustor
Содержание INTEGRA 3
Страница 76: ......