Section 10
Integra 3™ Troubleshooting
52
Research Instruments Ltd
10
Alarm
Condition
Code
Message
Priority
Fault
Description/
Alarm Criteria
Alarm
Actions
Solution
E01
HEATED
STAGE HEATING
FAILURE
Low
Heated Stage Plate
or Thermosafe Air
Heating System is
not able to heat.
Alarm activated
2 minutes after
power on if there
is less than 1.0°C
temperature rise
between 1 and
2 minutes after
power on. If the
temperature at 1
minute is already
within ±2.5°C from
the setpoint the
test is omitted.
Heated
Stage Plate
controller
power is set
to 0% until
the
Integra 3
is re-started.
Switch off the
Integra 3 and
verify that the
Heated Stage
Plate cables
and connector
are secure and
undamaged.
Possible heater
or controller
fault.
E02
THERMOSAFE
HEATING
FAILURE
Low
Thermosafe
Air Heating
System
controller
power is set
to 0% and fan
speed is set
to 0% until
the Integra 3
is re-started.
Switch off the
Integra 3 and
verify that the
Thermosafe
Air Heating
System cables
and connectors
are secure and
undamaged.
Possible heater
or controller
fault.
The following tables on pages 52 to 62 lists Alarm Condition Code
numbers, displayed Error Message, Priority, Fault Description, Alarm
Actions and a Solution to the fault.
Alarm Code Tables