After a quick initial blink, the status LED will blink steadily once the control box recognizes 
the  connection  to  the  laser  head.  If  the  initializing  LED  does  not  come  on  immediately, 
check the signal cable connection from the laser head to the back of the control box. 

 

The status indicator LED notifies the following: 

o

 

Starting- LED will blink slowly verifying laser emission. 

o

 

Acquiring- LED will blink rapidly signifying warming up and/or attempting to lock. 

o

 

Stable- LED stays on and does not blink meaning laser is locked in selected mode of 
operation. 

 

Ensure that no laser retroreflections enter the laser front bezel aperture during acquire. 

 

If the status LED fails to indicate the acquiring status after one minute, the laser tube is not 
igniting.  Switch the control box off in this case, wait several minutes and retry. 

 

 
BASIC TROUBLESHOOTING 

 
     Once the laser and control box have been properly connected, the stabilized laser system will 
operate without maintenance. Below are some guidelines if your system is not operating. 

 

No emission light above the key switch after key has been turned to the on position. 

o

 

Verify that the power cord is connected to the back of the control box as well as a 
verified power source.   

 

System fails to enter acquire status. 

o

 

Ensure that the remote interlock connector is well seated in the back of the control 
box. 

 

Status LED does not light. 

o

 

Check the control cable connection from the laser head to the back of the control 
box. 

 

No laser emission. 

o

 

Make sure shutter is open. 

o

 

Verify that  the high  voltage  cable  is  connected and  well  seated  to the back  of  the 
control box. 

o

 

If  laser  does  not  ignite  within  one  minute  switch  the  control  box  off,  wait  several 
minutes and retry. 

 

LED indicator moves in and out of lock status. 

o

 

Instability  is  usually  caused  by  retroreflections  entering  the  aperture  of  the  laser 
head.  Close the shutter and verify whether the system re-locks. This will determine 
whether  retroreflections  are  causing  instability  in the  system.  When the  shutter  is 
closed and the system remains locked, retroflections are causing the instability. 

 

The  specified  ambient  air  temperature  at  which  this  system  will  operate  is  15

C  -  30

C.   

Stability performance will be degraded if the system is operating at temperatures outside 
this range. For optimal performance, laser must be in a draft free environment where the 
ambient temperature remains constant and within the specified range (laser performance 
is optimal in free convection). 
 
 
 

Содержание 14309

Страница 1: ...GUIDE TO OPERATION Research Electro Optics Inc 5505 Airport Boulevard Boulder Colorado 80301 Phone 303 938 1960 Fax 303 447 3279 _______________________________________________________________________________________________________________________________ ...

Страница 2: ...ASER HEAD LABEL PLACEMENT DIAGRAM 6 FIGURE 2 POWER SUPPLY LABEL PLACEMENT DIAGRAM 6 INSTALLATION AND OPERATING INSTRUCTIONS 7 FIGURE 3 FRONT AND BACK PANEL OF CONTROL BOX 8 FIGURE 4 PROPERLY SEATED CONTROL CABLE 9 FIGURE 5 REMOVAL OF CONTROL CABLE 9 FIGURE 6 TWO POSITION SHUTTER ADJUSTMENT 10 FIGURE 7 FUSE REPLACEMENT 12 PRODUCT WARRANTY STATEMENT 12 EU DECLARATION OF CONFORMITY 14 ...

Страница 3: ...eturn needs to be made This package contains 1 Cylindrical Laser Head 1 Control Box 1 Control Cable 2 Keys 1 Remote Interlock Connector 1 115V Power Cord 1 230V Power Cord SAFETY INFORMATION This laser is safe to operate provided that the user complies with all safety warnings It is recommended that all personnel who will operate or be in the vicinity of the laser during operation read and be fami...

Страница 4: ...30990 LASER SYSTEM 633nm 5 0mW RANDOM 0 74 0 0026 13 30025 LASER SYSTEM 633nm 1 5mW POL 9 5 0 3 0 0026 5 63 TUNABLE LASER 633nm 4 0 mW POL 0 74 0 0026 13 TUNABLE LASER 612nm 2 5 mW POL 0 48 0 0026 7 02 TUNABLE LASER 604nm 0 5 mW POL 0 0 0026 0 TUNABLE LASER 594nm 0 6 mW POL 0 18 0 0026 3 68 TUNABLE LASER 543nm 0 3 mW POL 0 0 0026 0 31007 LASER SYSTEM 633nm 0 8mW POL 0 0 0026 0 31008 LASER SYSTEM 6...

Страница 5: ...ined by the National Center for Devices and Radiological Health CDRH Unauthorized modifications to the laser or power supply may cause irreversible damage to the system as well as result in possible hazardous radiation exposure These modifications will void all warranties COMPLIANCE Each Helium Neon laser system that is certified to be in compliance with the CDRH regulations is equipped with a key...

Страница 6: ...REF 176 6 Figure 1 Laser Head Label Placement Diagram Figure 2 Power Supply Label Placement Diagram ...

Страница 7: ...sure that it is recycled in a manner that protects human health and the environment For further information about where the user can drop off the waste equipment for recycling please email REO at henelaser reoinc com INSTALLATION AND OPERATING INSTRUCTIONS The cylindrical construction of the laser head allows for easy mounting in ring clamps V blocks or similar mounting hardware without affecting ...

Страница 8: ...de Selector Switch B Status LED Indicator C 2 Position Key Switch D High Voltage to Laser Head E Remote Interlock short to enable laser output F Control Signal to Laser Head Lemo G DB15 for Service Only H BNC Analog IN for Service Only I AC Input Fuse Assembly ...

Страница 9: ...peration at 115V or 230V Connect the control cable to the back of the control box Figure 3 F by aligning red dot on cable connector to red marking on port control box receptacle See Figure 5 and push to insert a mechanical click will indicate that it is properly seated Note when removing control cable make sure to pull the knurled portion backwards in order to release locking mechanism and keep it...

Страница 10: ...ure 5 Damage can occur to cable and or control box if not removed properly Insert the key into the front panel of the control box Figure 3 C Confirm that all connections between the laser and control box are well seated and or locked Plug the power cord into an electrical outlet that is confirmed operational Open the beam shutter located on the laser head at the output of the laser indicated by th...

Страница 11: ...ower cord is connected to the back of the control box as well as a verified power source System fails to enter acquire status o Ensure that the remote interlock connector is well seated in the back of the control box Status LED does not light o Check the control cable connection from the laser head to the back of the control box No laser emission o Make sure shutter is open o Verify that the high ...

Страница 12: ...rs made or Products or components replaced under warranty are warranted for the remaining unexpired duration of the original Warranty Period for such Products or components REO s standard repair charges shall be applicable forproducts returned for repair that are not covered under warranty Out of warranty repairs are warranted for ninety 90 days from date of shipment of the repaired Product The fo...

Страница 13: ...REO s designated facility REO will pay for shipment back to Buyer for Products repaired under warranty Note HeNe lasers are constructed from glass vacuum tubes and are very fragile REO strongly recommends taking appropriate precautions in preparing your laser for shipment Make sure the tube is well padded equivalent to 3 inch thick high density foam on all sides and insure that the power supply an...

Страница 14: ...art Number Description 32734 LASER SYSTEM 633NM STBLZD 1 5MW 32172 LASER SYSTEM 3 39UM 2 0MW POL 30990 LASER SYSTEM 633NM 5 0MW RANDOM 30025 LASER SYSTEM 633NM 1 5MW POL 9 5 31007 LASER SYSTEM 633NM 0 8MW POL 31008 LASER SYSTEM 633NM 0 5MW RAN 30968 LASER SYSTEM 543NM 0 5MW POL 30993 LASER SYSTEM 633NM 12 0MW POL 33361 LASER SYSTEM 543NM 1 5MW POL 30967 LASER SYSTEM 543NM 0 5MW RAN 30988 LASER SYS...

Отзывы: