38
XM-60 and XM-600 multi-axis calibrator
www.renishaw.com/xm60
Receiver alignment
•
To maintain the system accuracy the launch and the receiver must be
parallel to each other.
•
The software will check the parallelism on the completion of the fine
axis alignment.
•
If the parallelism tolerance is not met when selecting Capture, the
software will prompt the user to physically adjust the mounting of the
receiver. Rotate the receiver until the red border disappears. Do not
adjust the pitch and yaw adjustment of the launch unit.
NOTE:
It may be necessary to perform minor translation of the machine after
this step is complete.
NOTE:
Once set-up is complete and before capturing data, it is
recommended to perform an ambient light check. See the
CARTO Capture