49
www.renishaw.com/xm60
XM-60 and XM-600 multi-axis calibrator
Performance specifications
Linear
Axial range
0 m to 8 m
Mesurement range
0 m to 8 m
Accuracy
±0.5 ppm
(with environmental compensation)
Resolution
1 nm
Angular (pitch/yaw)
Axial range
0 m to 8 m
Measurement range
±500 µrad
Accuracy
±0.004A ±(0.5 + 0.11M) μrad
Resolution
0.03 μrad
A = displayed error reading
M = measured distance in metres
* The data stitch functionality in CARTO supports accurate readings for all six
degrees of freedom beyond 6 metres.
NOTE:
Accuracy values are reported to a statistical confidence of 95%
(k=2), they do not include the errors associated with the normalisation
of the readings to a material temperature of 20 °C.
Straightness *
Axial range
0 m to 6 m
Measurement range
±50 µm
±250 µm
Accuracy
±0.01A ±1 μm
±0.01A ±1.5 μm
Resolution
0.25 μm
Roll *
Axial range
0 m to 4 m
4 m to 6 m
Measurement range
±500 µrad
Accuracy
0 m to 4 m: ±0.01A ±6.3 μrad
4 m to 6 m: ±0.01A ±10.0 μrad
Resolution
0.12 μrad
All specifications assume the use of the XC-80 environmental compensator
and an environment with less than 1 °C change and a rate of change that
does not exceed 1 °C over a 20 minute period (0.05 °C/min).
Specifications do not include the effect of air turbulence (for example,
measurements at 4 m are obtained by averaging multiple sets of data).
The roll and straightness specification assume 1 second of averaging
of the signal.