14
Original instructions
12
Product specification
Set-up LED
Maximum speed
4 m/s
Acceleration
(readhead relative to scale)
< 200 m/s² in measuring direction
Moving force
(maximum force required to move the
readhead through the seals)
< 5 N
Vibration
(55 Hz to 2000 Hz)
Housing: < 300 m/s² to IEC 60068‑2‑6
Readhead: < 300 m/s² to IEC 60068‑2‑6
Shock 11 ms half-sine
< 300 m/s² IEC 60068‑2‑27
Operating temperature
0 °C to 50 °C
Storage temperature
–20 °C to 70 °C
Environment protection
IP53 when installed correctly, IP64 with air purge
Protection class III
Pollution degree II
Altitude 2000 m
EMC immunity
IEC 61800-5-2:2016
Electromagnetic immunity
requirement for safety related systems - Annex E, second
environment
Air purge requirements
Air supply pressure = 1 bar at encoder
At correct supply pressure the supplied air connection
fitting restricts the air flow rate to 2 l/min
Air quality: see
for details
Weight
0.27 kg + 2.0 kg/m
Measuring standard
Renishaw stainless-steel scale with single track absolute
encoding
Coefficient of thermal expansion
(at 20 °C)
10.1 ±0.2 μm/m/°C
Thermal datum
At centre position (encoder position of 0.5 × measuring
length)
Measuring lengths available (mm)
140, 240, 340, 440, 540, 640, 740, 840, 940, 1040, 1140,
1240, 1340, 1440, 1540, 1640, 1740, 1840, 2040, 2240,
2440, 2640, 2840, 3040
Accuracy grades
High grade: ≤ ±3 μm
Standard grade: ≤ ±5 μm
Resolution
1 nm, 10 nm, 50 nm
Absolute position protocols
BiSS Safety, Siemens DRIVE-CLiQ (with external
interface)
Encoder electrical connection
Cable connector M12 custom (customer configurable exit
direction)
Controller electrical connection
8‑way M12, 9‑way D‑type, flying lead
Cable length
Power supply
See
and
IMPORTANT:
Specifications are subject to the correct installation procedures as set out in this installation
guide. If in doubt, contact your local Renishaw representative.